4-1
IM WX103-01E
1
2
3
4
5
Index
Calibrator
4.1 Connecting the MX100 Data Acquisition Unit
Notes on Using the Calibration Software
If another user or another software program is connected to the MX100, drop the
connection before connecting with the Calibration Software.
If you change the module conguration before connecting with the Calibration
Software, carry out system conguration on the Logger (see section 2.2, “Conguring
the System”).
When calibrating, do not perform other operations. Use extra c
aution in the operation,
because the module may malfunction if you execute a wrong operation while
calibration is in progress.
Procedure
1. Start the Calibrator.
The Unit Information window opens.
2. In the
IP Address/Host Name box, enter the IP address or host name of the MX100
to be calibrated.
3. Click Connect.
Enter the host name or IP address of the MX100 to be calibrated.
Execute connection
When the connection is established, information about the unit’s module
conguration is displayed. If you click the Check button, the 7-segment LED on the
main module of the connected unit shows the text “—CALL—” owing from right to
left.
To disconnect the connected MX100
To turn on the 7-segment
LED on the connected
MX100 to check which unit
is the one connected
Information about the connected MX100
Modules attached to the
connected MX100 Data
Acquisition Unit
Chapter 4 Calibrator