The waveforms of the measurement channels and math channels are divided by the Mon.
Interval specified on the Acquisition screen of the Logger even when in the same group.
Click the unit number
Synchronized file
A file with .mxs extension
Unsynchronized file
A file with .mxd extension
Note
If you click the Yes button in the synchronization confirmation dialog box when a synchronized
file already exists, the message dialog box below opens. Click OK to load the synchronized file
and open the waveform display window.
Explanation
Loadable Files
• Data les recorded using the Logger of the MXLOGGER or the Integration Monitor of
the MX100 Standard Software (.mxd extension).
• Data les saved after performing synchronization using the Viewer (.mxs extension).
• Data les saved to the CF card using the MX100 (.mxd extens
ion).
• Data les saved after being joined on the Viewer (with the .mxc extension)
Display Range, File Size, and Number of Data Points of the Loadable Data
• 50 groups
• Maximum channels per group: 32
• Example: 2 GB
• Data points: 5 million
Note
• When loading a file of close to five million points, the error message “Insufficient Memory.
Close immediately.” appears. If this happens, set Total paging file size for all drives to 2 GB
or more. On PC, you can change the setting in the dialog box that opens by choosing from
the System Properties dialog box, Advanced > Settings under Performance > Advanced
> Change under Virtual memory.
• Please avoid loading multiple files containing close to five million data points. Doing so may
degrade the PC performance greatly.
3.1 Loading Data Files