4-17
IM WX104-01E
Displaying Data with the Data Viewer
From the menu bar
7. Click here(display files for linking).
5. Click here([Window]-[Link]).
6. The [Link] dialog box opens.
8. Select [Prev](previous file), [Next](next file), or [ALL](previous and next files).
9. Displays the linked files.
10. Displays the linked files in a different color.
The file extension .ldx is appended to the original file name
When linking and displaying files, make sure that the number of data points after linking
does not exceed 5000000.
In addition, if there is a period over which data does not exist such as when a power failure
occurs, data is counted as if the data is acquired at the given scan interval even during that
period. The scan interval and the maximum period for linking files are indicated below.
Interval Period
25 ms 36.4 hour
1/8 s 7.5 days
1 s 60.6 days
10 s 606.8 days
For example, if data is captured continuously at a 1/8 second scan interval and a power
failure occurs over a period of one week or longer, the data before and after the power
failure cannot be linked and displayed.
Saving the Link Settings File
Select [File] - [Save Display Setting As] to save the link settings file to the same directory
as the linked files.
The file name takes the form of the original file name with the file extension .ldx.
You can save the file by specifying the file name and the destination directory by
selecting [File] - [Save Display Setting].
4.5 Linking Files and Saving the Link Settings File