Error Message
No. Message Corrective Action
E3106 Data does not exist. Select another file.
E3107 Channel dose not exist. Select another file.
E3109 This name’s directory already
exits.
Overwriting is not possible when a directory with the same name exists.
Change the file name.
E3110 Can’t overwrite to file Open the file of the same name, close the other application, then retry the save.
E3111 Can't write to file. Please perform the operation on a disk with sufficient free space.
E3115 Too many data. Because the number of data for connection and display exceeds 5 million data
points, no file data is displayed. To check the file data that is not displayed, open
the corresponding file, then perform the connection and display procedure.
Warning Message
W3108 This file already exists.
Replace existing file?
W3115 Converting operation exceeded the limitation of MS Excel.
Continue converting?
W3115 Converting operation exceeded the limitation of Lotus 1-2-3.
Continue converting?
Configurator
Message
M6061 Data can’t be processed after the year 2038.
M6062 Any destroyed A/D converter exist. Any settings may be failed to store.
M6063 Receiving finished
M6064 Sending finished
Error Message
Message Corrective Action Reference Page
Communication error. Check the communication settings. 1-5
Connection timeout Traffic on the network may be congested.
Wait a few moments and try again.
Illegal user information Confirm whether the user name is correct. 5-24
Connection was refused. Check the communication settings. Check whether the
power to the DX/MV/CX is ON.
1-5
Connection is busy. Wait a few moments and try again.
Memory Error Exit other programs and restart, or restart the operating
system.
User level Error The currently logged on user ID is not authorized to
perform this action.
No. Message Corrective Action Reference Page
E6001 Failed to make file. Check the capacity of the directory.
E6002 Failed to open file. If the file still cannot be read after a second attempt, the
file may be damaged. Select another file.
E6003 Unreadable file Select another file. 5-1, 5-3
E6004 Communication impossible while
media in use
Send after media save is completed. 5-27
E6005 Now sampling & calculating.
Can't store settings.
Send after writing to the DX/MV/CX internal memory and
computation are complete.
5-27
E6006 Now sampling.
Can't store settings.
Send after writing to the DX/MV/CX internal memory is
complete.
5-27
E6007 Now calculating.
Can't store settings.
Send after computation is complete.
E6008 Now Controlling.
Can't store settings.
Send after control is finished.
9.1 Error Messages and Their Corrective Actions