Adjustment
Adjustment
Adjustment
(5) Set the Decimal value as follows:
a. If tenths (ex. XXX.X) set SW8 on.
b. If hundredths (ex. XX.XX) set SW7 on.
c. If thousandths (ex. X.XXX) set SW6 on.
(6) The values to check are shown in the table below.
(7) Set the voltage standard to the 0% value. Adjust the Zero Pot (top one) to read the minimum value
needed.
(8) Set the Voltage Standard to the 100% value. Adjust the Span Pot (bottom one) to read the
maximum value needed.
(9) Check all points in the table above and verify unit is within specification.