Yokogawa EJX210B Bedienungsanleitung Seite 78

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 101
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 77
<10. Parameter Summary>
10-1
IM 01C27C01-01EN
10. Parameter Summary
Table 10.1 Parameter
Object
ID
Attribute
ID
Label Description
Default
value
Handling
1.
UAPMO
block
1 Version Revision Indicates the application revision of EJX This revision
when the application software is downloaded.
--- R
10 Static Revision Indicates the revision level of the xed parameters of
UAP Used, for example, to check whether parameters
have been change.
0 R
64 Identication
Number
Indicates the vender ID, model ID, and revision of the
device.
--- R
65 CTS Version Indicates the version of the communication stack test
system (CTS).
0 R
66 ITS Version Indicates the version of the inter operability test system
(ITS).
0 R
67 Diagnostic Status Indicates the diagnostic results of the device based on
the NAMUR NE107
*1
model.
Setting Enable diagnostic status conguration in UAP
Option to Enable allows turning OFF and ON the
display of the diagnostic results for each summary,
and changing Categorize For Categorize at the time of
shipment, refer to tables 8.3 and 10.2.
--- R
68 UAP Option Allows setting the Diagnostic Status and write protection
of UAP.
Software write protect
1: On, 0: Off
Enable hardware write protect
1: Enable, 0: Disable (default)
Enable diagnostic status conguration
1: Enable, 0: Disable (default)
The following table shows the relationship between the
hardware write protection and software write protection.
Enable
hardware
write
protect
Hardware
write
protect
Software
write
protect
Write
protect
Disable Off or On Off No
(Writable)
Disable Off or On On Yes
(Unwritable)
Enable Off Off or On No
(Writable)
Enable On Off or On Yes
(Unwritable)
0 W
69 Diagnostic
Switch
Allows setting On/Off for each summary of Diagnostic
Status when Enable diagnostic status conguration in
UAP Option is set to Enable.
On W
70 Diagnostic
Conguration
Allows Categorize for each summary of Diagnostic
Status when Enable diagnostic status conguration in
UAP Option is set to Enable.
Refer to Table
10.2.
W
102 Diagnostic Status
Detail[2]
Detailed information on Diagnostic Status. For
Categorize at the time of shipment, refer to Table 10.2.
Refer to Table
10.2.
R
103 Energy Left Indicates the number of days of remaining battery life
assuming ambient temperature condition as 23 degrees
Celsius.
The unit is day.
--- R
104 Reset Energy Left Initialize the remaining battery life (Energy Left) and
reset it as new batteries.
Perform the battery replacement.
0 = Continue (Cancel)
1 = Reset (Initialization)
0 (reading value
is always 0)
W
Seitenansicht 77