<3. INSTALLATION>
3-5
IM 01E25D11-01EN
(2) Carrying Flowtube
F0309.ai
90 degrees or less
90 degrees or less
Eye bolt
Eye plate
Figure 3.3.1 Carrying Flowtube
CAUTION
When carrying the owtube, use the lifting
rings (eye bolts or eye plates).
To assure safety, keep lifting angle less than
90 degrees as shown in Figure 3.3.1
(3) Positioning Flowtube
Bring in the owtube, place it and use a jack to
adjust its position.
IMPORTANT
Apply the jack to the anges of the owtube.
In addition, adjust any misalignment when the
owtube is brought in, as the jack can adjust
vertical position, but not horizontal one.
(4) Applying Gasket and optional Grounding
Rings
Gasket:
Necessary gaskets for piping connection are
as below depending on the choice of grounding
rings.
Use sheet gaskets designed to t for ange
standard. The GF type-1 gaskets by JIS G
3443-2 should be used for the JIS F12 ange
models in sizes 1100 mm (44 in.) and above. In
this case, the gasket groove is required on the
customer pipe anges.
The thickness of gasket should be 2mm
(0.08in.) to 5mm (0.2in.) for sizes up to 1000
mm (40 in.), and 5 mm (0.2 in.) or more
for larger sizes. The type of gasket should
be a kind of soft rubber, or its equivalent in
hardness.
1. Standard (no grounding rings)
Gaskets A supplied by customer
Customer
pipe
Customer
pipe
A A
F0310.ai
Figure 3.3.2 Installation without Grounding
Rings
When using the GF type-1 gaskets, the
gasket groove is required as mentioned
above.
2. With optional grounding rings (code GR1)
Optional grounding rings
Gaskets A and B supplied by customer
Customer
pipe
Customer
pipe
A
B
B
A
F0311.ai
Figure 3.3.3 Installation with Grounding Rings
GR1
It is recommended to use the same gasket
for A and B. Both gaskets A and B are to be
supplied by customer.
When polyurethane or natural soft rubber
lining, the gaskets B are not necessary.
3. With optional grounding rings (code GR2)
Customer
pipe
Optional grounding rings
with integrated gaskets B included
Gaskets A supplied by customer
Customer
pipe
F0312.ai
A A
B
B
Figure 3.3.4 Installation with Grounding Rings
GR2
This is available for the models with
process connection JIS F12 (JIS 75M) in
sizes 1100 mm (44 in.) and above. Only
gaskets A are to be supplied by customer.
The grounding rings are installed and
locked onto the owtube with gaskets B
when shipped from factory.
Mounting Procedure (no Grounding Rings):
Connect the owtube’s ange and the
customer’s pipe which contact processuid by
some wire supplied by customer.