<6. Setup>
6-5
IM 12A01S02-01E 1st Edition : May 10, 2013-00
Table 6.3 pH/ORP common settings (numbers in parentheses indicate related sections of FLXA21
User’s Manual (IM 12A01A02-01E))
Setting Item Available Option Initial Value Remarks
pH+ORP measurement pH, ORP, pH+ORP,
pH+rH, rH
pH+ORP
TimeAxisRange(Max) 1 to 10 min. 5 min. Timeline displayed
on the calibration
screen.
pH
calibration
settings
Zero and
Slope units
Zero unit mV, pH mV
Slope unit %, mV/pH %
Select buffer set NIST/DIN19266,
DIN19267,US,
Programmable
NIST/DIN19266
Setup 0.00 to 20.00 pH (Table 6.4) Appears when
selecting
“Programmable”.
Temperature difference 1.0 to 5.0 degC
-1.8 to 9.0 degF
2.0 degC
3.6 degF
Setsthemaximum
temperature offset
between sensors for
multiple calibration.
Limits Zero high 0.00 to 500.0 mV
0.00 to 8.45 pH
120.0 mV
2.03 pH
Zero low -500.0 to 0.00 mV
-8.45 to 0.00 pH
-120.0 mV
2.03 pH
Slope high 100.0 to 110.0 %
59.16 to 65.08 mV/pH
110.0 %
65.08 mV/pH
Slope low 70.0to100.0%
41.41 to 59.16 mV/pH
70.00%
41.41 mV/pH
Step Range 0.01 to 1.00 pH 0.01 pH
Stabilization
time
2 to 30 s 10 s
ORP
calibration
settings
Limits Zero high 0.00 to 500.0 mV 120.0 mV
Zero low -500.0 to 0.00 mV -120.0 mV
Slope high 100.0 to 110.0 % 110.0 %
Slope low 70.0to100.0% 70.00%
Step Range 1 to 100 mV 1 mV
Stabilization
time
2 to 30 s 10 s
rH
calibration
settings
Limits Zero high 0.00 to 500.0 mV 120.0 mV
Zero low -500.0 to 0.00 mV -120.0 mV
Slope high 100.0 to 110.0 % 110.0 %
Slope low 70.0to100.0% 70.00%
Step Range 0.01 to 10.00 rH 0.05 rH
Stabilization time
2 to 30 s 10 s
Impedance
settings
Impedance 1 high limit 1.000 kOhm to
1.000 MOhm
(> Impedance 1 low
limit)
200.0 kOhm When Impedance 1
is “low*.”
In the case of
“high,” use the
value of FINE limit
for impedance
1 in SENCOM
diagnostic criteria.
Impedance 1 low limit 1.000 kOhm to
1.000 MOhm
(< Impedance 1 high
limit)
1.000 kOhm
Impedance 2 high limit 1.000 kOhm to
1.000 MOhm
(> Impedance 2 low
limit)
200.0 kOhm When Impedance 2
is “low*.”
In the case of
“high,” use the
value of FINE limit
for impedance
2 in SENCOM
diagnostic criteria.
Impedance 2 low limit 1.000 kOhm to
1.000 MOhm
(< Impedance 2 high
limit)
1.000 kOhm