Yokogawa FieldMate Bedienungsanleitung Seite 48

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 84
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 47
<6 OPERATION>
6-16
IM 01R01A11-01EN
1
2
F060204.ai
Figure 6.2.4 “No Flow ConrmationDialog
Table 6.2.4 “No Flow Conrmation” Dialog
Item
No.
Item Name Description
1 OK Runs “Circuitcheck and “Device Status” check.
2 Cancel
Stops runningCircuit” check and “Device Status” check, and displays
“Main Screen”.
4) Click OK (1)“ after conrming that the owtube is lled with uid and the uid velocity is
completely zero. “Circuit” check and “Device Status” check are run. During these checks,
the status bar and remaining time are displayed. See Figure 6.2.5. Regarding the execution
time, see Table 6.2.5.
F060205.ai
Figure 6.2.5 “During ”Circuit” Check and “Device StatusCheck” Screen
Table 6.2.5 Execution Time [No Flow]
No. Optional code Execution time
1 Without ELC 630 s
2 With ELC (DC Noise Cut Circuit) 120 s
Seitenansicht 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 83 84

Kommentare zu diesen Handbüchern

Keine Kommentare