<9. General Specications>
9-5
IM 01C25B01-01E
Physical Specications
Wetted Parts Materials
Diaphragm, Cover Flange, Process
Connector, Capsule Gasket, and Vent/Drain
Plug
Refer to “Model and Sufx Code.”
Process Connector Gasket/o-ring
[EJ110, EJ120, EJ430, EJ310]
PTFE Teon
Fluorinated rubber for Option code /N2 and /N3
[EJ130]
Fluorinated rubber (O-ring)
[EJ440]
Fluorinated rubber (O-ring) for C capsule
Glass reinforced Teon (gasket) for D capsule
Non-wetted Parts Materials
Bolting
ASTM-B7 carbon steel, 316L SST stainless
steel, or ASTM grade 660 stainless steel
Housing
Low copper cast aluminum alloy with
polyurethane paint or ASTM CF-8M stainless
steel
Degrees of Protection
IP66/IP67, NEMA TYPE 4X
Cover O-rings
Buna-N, uoro-rubber (option)
Nameplate and tag
316SST (including /N4 wired tag)
Fill Fluid
Silicone, Fluorinated oil (option)
Weight
[Installation code 7, 8, and 9 and wetted parts
material code S for EJ110, EJ310, and
EJ430 (Except for measurement span code
F)]
2.8 kg (6.2 lb) without integral indicator,
mounting bracket, and process connector.
[Installation code 7, 8, and 9 for EJ120]
3.7 kg (8.2 lb) without integral indicator,
mounting bracket, and process connector.
[Installation code 7, 8, and 9 for EJ130]
6.8 kg (14.3 lb) without integral indicator,
mounting bracket, and process connector.
[Installation code 7, 8, and 9 for EJ440]
4.9 kg (10.8 lb) without integral indicator,
mounting bracket, and process connector.
Add 1.5 kg(3.3lb) for amplier housing code 2.
Connections
Refer to “Model and Sufx Code.”
Process Connection of Cover Flange (models
except for EJ440-D): IEC61518