6-17
IM DR231-11E
Commands
6
6.6 Control and Execution
Command
PS Starts or stops recording.
Mode Operation mode
Setting PSp1<terminator>
p1 0 Recording starts.
1 Recording stops.
MP Starts or stops manual printing.
Mode Operation mode
Setting MPp1<terminator>
p1 0 Manual printing starts.
1 Manual printing stops.
Example Start manual printing.
MP0
LS Starts or stops list printing.
Mode Operation mode
Setting LSp1<terminator>
p1 0 List printing starts.
1 List printing stops.
Example Start list printing.
LS0
HD Starts or stops header printing.
Mode Operation mode
Setting HDp1<terminator>
p1 0 Header printing starts.
1 Header printing stops.
Example Start header printing.
HD0
SU Starts or stops setup list printing.
Mode Operation mode
Setting SUp1<terminator>
p1 0 Setup list printing starts.
1 Setup list printing stops.
Example Start setup list printing.
SU0
MS Starts message printing.
Mode Operation mode
Setting MSp1<terminator>
p1 01 to 20 (message number)
Example Start printing message No. 5.
MS05
AK Confirms the current alarm status.
Mode Operation mode
Setting AKp1<terminator>
p1 0 Confirms the current alarm status.
Example Confirm the current alarm status.
AK0
AR Resets an alarm.
Mode Operation mode
Setting ARp1<terminator>
p1 0 The alarm is reset.
Example Reset the alarm.
AR0
IR Resets the timer.
Mode Operation mode
Setting IRp1<terminator>
p1 0 The timer is reset.
Example Reset the timer.
IR0
AC Clears the alarm buffer.
Mode Operation mode
Setting ACp1<terminator>
p1 0 The alarm buffer is cleared.
Example Clear the alarm buffer.
AC0
MC Clears the message buffer.
Mode Operation mode
Setting MCp1<terminator>
p1 0 The message buffer is cleared.
Example Clear the message buffer.
MC0
EX Computation start/stop, restart of
computed data after they are cleared,
and release of statuses after
completing measurement.
Mode Operation mode
Setting EXp1<terminator>
p1 Computation start/stop, restart of computed
data after they are cleared, and execution of
releasing statuses after completing
measurement.
0: Computation start
1: Computation stop
2: Restart of computed data after they are
cleared
3: Computed data clear
4: Release of statuses after completing
measurement
Example Start the computation.
EX0
Comments This command is effective only with the optional
computation functions or a floppy disk drive
function or with a pulse module installed.
If MATH is set for level action in the event/action,
a computation start/stop/start after clearing the
computed data cannot be done using EX
command.
This command is not executable during the
saving/reading of setup data.
MW Stores the measured data/computed
data on a RAM disk.
Mode Operation mode
Setting MWp1, p2, p3, p4, p5<terminator>
p1 Retention mode
DIRECT Data storage starts immediately
after sending an MW command.
TRIGER Data storage starts after a trigger
signal is generated.
REPEAT Specified data are repeated for
storage each time the trigger
signal is generated.
p2 File name (up to eight characters for other
than REPEAT, up to five characters for
REPEAT)
p3 Retention interval
INTVL Data are stored at measurement
intervals.
1 min/2 min/5 min/10 min
Data are stored at specified
intervals.
6.6 Control and Execution Command