Yokogawa DR240 Bedienungsanleitung Seite 46

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 121
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 45
5-3
IM DR231-11E
Command Format
5
5.2 Command Syntax
In this manual, each command is explained as shown below.
SD Set date & time.
Mode Operation mode
Set SDp1, p2<terminator>
p1 year, month, day
p2 hour, min., sec.
Ex. Set the clock inside the DA to July 1, ’96, 13:00:00
SD96/07/01, 13:00:00
Comments • The formats of p1 and p2 are fixed to 8 characters.
Set them in the following manner.
p1=YY/MM/DD
p2=HH:MM:SS
Command function
Command
Effective command mode
Command syntax
Parameter description
Example of setting
Explanation of the way to
use and remarks in more detail
Mode
There are the following three modes in the DR130/DR230/DR240.
Operation (measurement) mode
Mode to perform normal operation (measurement).
Setup mode
Mode to set the basic specifications for the DR130/DR230/DR240.
To confirm the settings, execute the XESTORE command. See the description for the XE command.
A/D calibration mode
Mode to execute calibration of the A/D module.
Each mode is selected by the DS command. For details, see DS Command on page 6-20.
Also in each mode, commands and parameters that can be set and parameters that can be output
differ. See the description for each command.
Number of Channels
The configurable numbers of measurement and computation channels vary, from recorder to recorder,
between model DR130, stand-alone models of the DR230/240, and expandable models of the DR230/
240, as shown below:
Type of DR Recorder Highest Measurement-channel Highest Computation-channel
Number Number
DR130 020 A30
Stand-alone models 030 A30
of DR230/240
Expandable models 560 A60
of DR230/240
Seitenansicht 45