<6. PARAMETER DESCRIPTION>
6-28
IM 01E30D01-01EN
Output Example 2
Thehighalarm(H)issetto80%ormoreoftheow
rate span only without low alarm (L) setting; and the
H/L alarm hysteresis width is set to 0%.
Settings are:
G10: Low Alarm=-105%
A setting value of -105% indicates that the
alarm is disabled.
G11: High Alarm=80%
G12:H/LAlarmHys=0%
0 Time (t)
Instantaneous flow rate
50
80
100
[%]
Select “H/L Alarm” for F10: DO Function
Select “Closed (On) Act” for F11: DO Active Mode
Select “Yes” for G28: Alm-H/L
High Alarm
F0606.ai
DO Open Closed Open
[G12:H/LAlarmHys] Setting of upper/lower alarm
value hysteresis width
This parameter sets the hysteresis width for upper
and lower alarm value, using a % value of the
maximum span.
Output Example
The hysteresis width is set to 5%.
Settings are:
G10: Low Alarm=30%
G11: High Alarm=80%
G12:H/LAlarmHys=5%
0
Time (t)
35%
Instantaneous flow rate
30
50
80
100
[%]
Select “H/L Alarm(O)” for F10: DO Function
Select “Closed (On) Act” for F11: DO Active Mode
Select “Yes” for G28: Alm-H/L
High Alarm
Low Alarm
F0607.ai
75%
DO Open Closed Open OpenClosed
[G20: 4-20mA System Alarm] Display of the
current output during a system alarm occurrence
This parameter displays the current output during a
system alarm occurrence.
Display Function
3.2 mA(see IM) Fixed at 3.2 mA or less
21.6 mA(see IM) Fixed at 21.6 mA or more
The default parameter is set to the following table at
the factory.
Standard Optional code C1
21.6 mA(see IM) 3.2 mA(see IM)
NOTE
Thedirectionofcurrentoutputcanbemodied
when a system alarm occurs. Refer to Section
10.3.1, “Setting of Burnout Switch.