<4. WIRING>
4-14
IM 01E30D01-01EN
Table 4.6.6 TIIS Explosion Proof Type Wiring Examples (Simultaneous Current-Pulse Output)
Connection Description
Simultaneous
Current-Pulse Output
When simultaneous output of current and pulse output are used, no communication is possible
in some cases. Refer to following example 1 to 3.
Example 1
In this case,
Communication is
possible (up to a
distance of 2 km
when a CEV cable
is used) and when a
two-wire or three-
wire shielded cable
is used.
No communication is possible
when a shielded cable is not used.
However, simultaneous current-pulse output is possible.
Shielded Cable
Distributor, etc.
R3
R
Shielded Cable
For the shielded cables in this example,
use two-wire shielded cables
separately for SUPPLY and DO.
E1
AXR Terminal
E3:30 V DC max
*8
*7
*6
*3
*3: This supply voltage requires a power source
with a maximum output current of no less than E3/R.
Electric
counter
R3[Ω]:Load Resistance
JIS Class A
SUPPLY
DO
0.0236 x (R1 + R2+ R3) + 14.7 ≤ E1 [V] ≤ 42
*10
In the case of R3=250[Ω]
0.0236 x (R1 + R2) + 20.6 ≤ E1 [V] ≤ 42
*10
Cable Resistance:R1[Ω] *9
Cable Resistance:R2[Ω] *9
+
+
+
+
Hazardous area Non-hazardous area
Example 2
In this case,
Communication is
possible (up to a
distance of 2 km
when a CEV cable
is used) and when a
one-wire or two-wire
shielded cable is
used.
Shielded Cable
Distributor, etc.
R3
R
Shielded Cable
For the shielded cables in this example,
use two-wire shielded cables
separately for SUPPLY and DO.
E1
AXR Terminal
E3
:
30 V DC max
0.0236 x (R1 + R2+ R3) + 14.7 ≤ E1 [V] ≤ 42
*10
In the case of R3=250[Ω]
0.0236 x (R1 + R2) + 20.6 ≤ E1 [V] ≤ 42
*10
*7
*6
*4: This supply voltage requires a power source
with a maximum output current of no less than E3/R.
R3[Ω]:Load Resistance
*4
JIS Class A
*8
SUPPLY
DO
Cable Resistance:R1[Ω] *9
Cable Resistance:R2[Ω] *9
Hazardous area Non-hazardous area
Electric
counter
+
+
+
+
Example 3
In this case,
No communication is
possible when a four-
wire cable is used.
R
250Ω
E5
0.0236×﴾R1
+
R2﴿
+
20.6E5
[
V
]
30
*7
AXR Terminal
Electric
counter
*6
*5
JIS Class A
*8
SUPPLY
DO
Cable Resistance
:
R1[Ω] *9
Cable Resistance
:
R2[Ω] *9
Recorder
or other
instrument
+
+
+
Hazardous area
Non-hazardous area
*5: This supply voltage requires a
power source with a maximum
output current of no less than (E5/R+0.0236).
The range of load
resistance R for the
pulse output
T
herangeofloadresistanceRforthepulseoutputmustbasicallybe1and2W.Theload
resistance should be selected by calculation as shown below when proper transmission is
impossible due to the length of cable or frequency of pulse output.
P (mW) =
…………………………
(2)
R (kΩ)
E
2
(V)
R (kΩ)
(1)
120
E (V)
C (μF) × f (kHz)
0.1
E: Supply voltage (V)
f: Frequency of pulse output (kHz)
R
:Valueofloadresistance(kΩ)
C:CablecapacitanceF)
P:
Electrical power of the load resistance (mW)
Note: C 0.1F/km)forCEVcable
Note:
ThecommunicationispossiblethoughitmightnotmeetapartoftheHARTcommunicationspecicationdependingonuseconditions.
W
henusingcurrentandpulseoutputsimultanously,theHARTcommunicationmaybeinuencedbynoisecomparinganalogoutputonly.
Seitenansicht 37