IM 1E10C1-E
11-10
11. PARAMETER LIST
Tl set value 0 to 999999 0 W Sets forward direction totalization preset value. E05
Total set (00) Inhibit Inhibit W Inhibits or enables forward direction totalization and reverse E04
(01) Enable direction totalization from being preset and reset, respectively.
Total switch 0 to 999999 0 W Sets switching level when using status output function as E06
totalization switch function.
Reverse total 0 to 999999 0 W Displays and resets reverse direction totalizaed flow rate. A31
Diff. total –999999 to 999999 0 R Displays differential totalized flow rate between forward and A32
reverse.
(Diff. total = Totl – Reverse total)
Analog low cut 0 to 10% 0 W Sets low cut width of current output. G01
Analog low lmt –20 to 100% –20 W Sets low limit of current output. G02
Analog hi lmt 0 to 120% 120 W Sets high limit of current output. G03
PV AO Alrm typ Hi R Displays burn out direction of current output.
Analog alarm out (00) 2.4 mA or less 21.6 mA W Selects current output during alarm occurrentce. b11
(01) 4.0 mA or more
(02) Hold
(03) 21.6 mA or more
Poll addr 0 to 15 0 W Sets polling address when multidrop mode.
Num req preams 5 5 R Displays number of request preambles.
Burst mode Off Off W Selects the mode for the burst mode functionality.
On
Burst option PV PV W Selects sending items (instantaneous flow rate, output in %,
% range/current totalization value and/or current output) when burst mode.
Process vars/crnt
Output function (00) Pulse out Pulse out W Selects status output functions. b10
(01) Alarm out
(02) Bi direction
(03) Auto 2 ranges
(04) Low alarm
(05) Total switch
Reverse span 0.0001 to 30000 1.0000 W Sets span for reverse direction. b30
Bi direction hys 0 to 10% 2 W Sets hysteresis when change in flow direction. b31
Forward span2 0.0001 to 30000 1.0000 W Sets the second span for forward direction. b33
Auto range hys 0 to 15% 10 W Sets hysteresis when change in Auto 2 ranges. b34
Low alarm –10 to 110% –10 W Sets low limit alarm. b36
Low alarm hys 0 to 10% 5 W Sets hysteresis for low limit alarm. b37
Pulse unit (00) n UNIT/P PULSE/s W Selects pulse rate unit. F01
(01) u UNIT/P (“UNIT” in the selection on the left represents the “Base vol
(02) m UNIT/P unit” selected.)
(03) UNIT/P
(04) k UNIT/P
(05) M UNIT/P
(06) PULSE/s
Name Data Range, Units
Default
Value
R/W Description
Indicator
of SE