<5. Error Information>
5-1
IM 12A01S02-01E 1st Edition : May 10, 2013-00
5. Error Information
When an error occurs, the color of [Error information] on the toolbar changes and an icon starts to
ash.
Clicking on [Error information] displays the error that is occurring in the connected sensor.
The error information screen displays the error that is occurring in any connected sensors, and
any error or warning for the SPS24 itself. Errors are more critical than warnings.
Clicking on the error or warning information displays the actions required.
l [History]
Clicking on [History] on the error information screen displays the error history screen.
The error history screen shows sensor-dependent errors and general errors separately.
Wheneverclickinganyconnectedsensorbuttonsintheconnectioninformationeld,thesensor-
dependenterroreldswitchesinformationfortheselectedsensor.
Whenthetotalofgeneralerrorsandsensorerrorsexceeds100,olderinformationwillbedeleted.
“Type” in the display item indicates F (Fault) or W (Warning).
[Clear all] clears all error history.
l Connected Sensor Button
Action is required when the connected sensor button displays the following status.
The stored setting for the connected sensor is corrupted.
Execute[Initializesensor]from[Sensormanagement]–[Sensorsetup].
The connected sensor is not supported.
Update SPS24 to support the connected sensor.
Beforeupdating,itisrecommendedthatyouexecute[Export]forthedatabase
viewer and the setup.
This indication does not change the display color of [Error information] on the
toolbar.
Neither is it registered in the history.