(III) Specications regarding ll uid code 1, 2, and 4 (for EJA118E, EJX118A, EJA438E, and EJX438A)
Please use the following additional contents for each item on the section 10, when ll uid code 1, 2, or 4 is specied.
8.1 Standard Specications (for EJA118E, EJX118A, EJA438E, and EJX438A)
Working Pressure Limits (Page 8-2): See below table.
Fill uid Working pressure
High temp. and high vacuum use
High vacuum use
0.013 kPa abs (0.0019 psi abs) to ange rating pressure
For atmospheric pressure or below, see gure 8.a, 8.b, and 8.c.
-50 0 50 100 150 200 250 300 350
2.7(0.38)
100(14.5)
0.1(0.014)
0.013(0.0019)
0.01(0.0014)
1(0.14)
10(1.4)
Temperature (°C)
Process temperature
for fill fluid code 1
Flange max.
working
pressure
Atmospheric
pressure
Transmitter ambient
temperature range
(for fill fluid code 1)
*1
Working
pressure
kPa abs
(psi abs)
-50 0 50 100 150 200 250 300 350
2.7(0.38)
100(14.5)
0.1(0.014)
0.013(0.0019)
0.01(0.0014)
1(0.14)
10(1.4)
Temperature (°C)
Process temperature
for fill fluid code 2
Flange max.
working
pressure
Atmospheric
pressure
Transmitter ambient
temperature range
(for fill fluid code 2)
*1
Working
pressure
kPa abs
(psi abs)
-50 0 50 100 150 200 250 300 350
2.7(0.38)
100(14.5)
0.1(0.014)
0.013(0.0019)
0.01(0.0014)
1(0.14)
10(1.4)
Temperature (°C)
Process temperature
for fill fluid code 4
Flange max.
working
pressure
Atmospheric
pressure
Transmitter ambient
temperature range
(for fill fluid code 4)
*1
Working
pressure
kPa abs
(psi abs)
Figure 8.a. Working Pressure and
Process Temperature
(Fill uid: silicone oil for high
temp. and high vacuum use)
Figure 8.b. Working Pressure and
Process Temperature
(Fill uid: silicone oil for high
temp. and high vacuum use)
Figure 8.c Working Pressure and
Process Temperature
(Fill uid: silicone oil for high
vacuum use)
*1: The upper ambient temperature limit is 50°(122°F) in the following combinations.
Process Temperature Limits (Page 8-2): See below table.
Code Process temperature Ambient temperature Working pressure Specic gravity
Silicone oil
(high temp. and
high vacuum use)
1
–10 to 250°C
*1
(14 to 482°F)
–10 to 60°C
*2
(14 to 140°F)
0.013 kPa abs
(0.0019 psi abs)
to ange rating
pressure
1.07
Silicone oil
(high temp. and
high vacuum use)
2
10 to 310°C
(50 to 590°F)
10 to 60°C
*2
(50 to 140°F)
1.09
Silicone oil
(high vacuum use)
4
–10 to 100°C
(14 to 212°F)
–10 to 60°C
*2
(14 to 140°F)
1.07
*1: In case of wetted parts material code TW (Tantalum), process temperature limit is up to 200°C (392°F).
*2: The upper ambient temperature limit is 50°C(12F) in the following combinations.
Process connection style code Process connection size code
W (Flush type) 2 (2-inch) or 8 (1 1/2-inch)
E (Extension type) 3 (3-inch)