Yokogawa EJA440E Bedienungsanleitung Seite 39

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 41
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 38
<Appendix 1. Safety Instrumented Systems Installation>
A1-2
IM 01C25T03-01E
Table A1.2 Proof Testing
Testing method Tools required Expected outcome Remarks
Functional test:
1. Follow all Management of
Change procedures to bypass
logic solvers if necessary.
2. Execute HART/BRAIN
command to send value to
high alarm (21.5 mA) and
verify that current has reached
this level.
3. Execute HART/BRAIN
command to send value to
low alarm (3.6 mA) and verify
that current has reached this
level.
4. Restore logic solvers
operation and verify.
Handheld terminal Proof Test Coverage =52% The output needs to
be monitored to assure
that the transmitter
communicates the
correct signal.
Perform three point calibration
along with the functional test listed
above.
Handheld terminal
Calibrated pressure
source
Proof Test Coverage =99%
A1.2.6 Repair and Replacement
If repair is to be performed with the process online
the transmitter will need to be bypassed during the
repair. The user should setup appropriate bypass
procedures.
In the unlikely event that the transmitter has a
failure, the failures that are detected should be
reported to Yokogawa.
When replacing the transmitter, the procedure in the
installation manual should be followed.
The personnel performing the repair or replacement
of the transmitter should have a sufcient skill level.
A1.2.7 Startup Time
The transmitter generates a valid signal within 1
second of power-on startup.
A1.2.8 Firmware Update
In case rmware updates are required, they
will be performed at factory. The replacement
responsibilities are then in place. The user will not
be required to perform any rmware updates.
A1.2.9 Reliability Data
A detailed Failure Mode, Effects, and Diagnostics
Analysis (FMEDA) report is available from
Yokogawa with all failure rates and failure modes.
The transmitter is certied up to SIL2 for use
in a simplex (1oo1) conguration, depending
on the PFDavg calculation of the entire Safety
Instrumented Function.
The development process of the transmitter is
certied up to SIL3, allowing redundant use of
the transmitter up to this Safety Integrity Level,
depending the PFDavg calculation of the entire
Safety Instrumented Function.
When using the transmitter in a redundant
conguration, the use of a common cause factor
-factor) of 2% is suggested. (However, if the
redundant transmitters share an impulse line or if
clogging of the separate impulse lines is likely, a
common cause factor of 10% is suggested.)
Note that the failure rates of the impulse lines need
to be accounted for in the PFDavg calculation.
A1.2.10 Lifetime Limits
The expected lifetime of the transmitter is 50
years. The reliability data listed the FMEDA report
is only valid for this period. The failure rates of the
transmitter may increase sometime after this period.
Reliability calculations based on the data listed
in the FMEDA report for the transmitter lifetimes
beyond 50 years may yield results that are too
optimistic, i.e. the calculated Safety Integrity Level
will not be achieved.
A1.2.11 Environmental Limits
The environmental limits of the transmitter are
specied in the user’s manual IM 01C25.
A1.2.12 Application Limits
The application limits of the transmitters are
specied in the user’s manual IM 01C25. If the
transmitter is used outside of the application limits,
the reliability data listed in A1.2.9 becomes invalid.
Seitenansicht 38
1 2 ... 34 35 36 37 38 39 40 41

Kommentare zu diesen Handbüchern

Keine Kommentare