3.16 Attributes of Batch Group Number
Components
(Can only be set with a Version 4.01 screen and a Version 4.11 screen)
The following restrictions apply to the attribute settings of batch group number
components.
[Visible] can only be set when [Depend ID] is set to [None].
The MultiBatch number specied in the settings le is the maximum value for [Batch
group no.].
If the Batch function is Off or On in the settings le, a batch group number component
cannot be created. For existing batch group number components, no attributes other
than X, Y, Width, and Height can be set.
List of Settings
Refer to Section 3.3 for attributes without explanations in the list of settings.
Attribute Set value/choice
(Underlined items are initial
set values)
Description, conditions
X DX1000: 0 to 318
DX2000: 0 to 638
This is the X-coordinate of the left side of the component.
X = 0 for X < 0
X = (Max.
width
*1
– Width) for (X + Width) > Max. width
*1
Y DX1000: 24(0) to 238
DX2000: 40(0) to 478
( ) = Status area set to No
display
This is the Y-coordinate of the right side of the component.
When Y < min. Y
*2
, Y = min. Y
*2
When (Y + size) > (min. Y
*2
+ max. height
*3
), Y = (min. Y
*2
+ max.
height
*3
- size)
*1 DX1000 max. width = 320, DX2000 max. width = 640
*2 DX1000 min. Y = 24(0), DX2000 min. Y = 40(0), ( ) = Status area set to No display
*3 DX1000 max. height = 216(240), DX2000 max. height = 440(80), ( ) = Status area set to No display