IM 12D08N05-01E4th EditionUser’sManualModel SC450G [Style: S2]Conductivity / ResistivityConverter
2 IM 12D08N05-01E2. GENERAL SPECIFICATIONS OF EXAxt SC450GA) Inputs specifications : Two or four electrodes measurement with square wave excitation,
3IM 12D08N05-01E2 GENERALSPECIFICATIONS Hold : ContactcanbeusedtosignaltheHOLDsituation. Failsafe : ContactS4isprogrammedasafai
4 IM 12D08N05-01EO) Safety and EMC conforming standards , Safety : EN61010-1 CSAC22.2No.61010-1 UL61010-1 FM3611ClassI,
5IM 12D08N05-01E3 INSTALLATIONANDWIRING3-1. Installation and dimensions3-1-1. Installation siteThe EXAxt 450 converter is weatherproof and can be i
6 IM 12D08N05-01Emin.185 (7.25")min.195(7.75")138(5.43")138(5.43")M6M6M5138(5.43")+1 0138(5.43")+1 0Figure 3-2. Opt
7IM 12D08N05-01E3 INSTALLATIONANDWIRING3-2. Wiring3-2-1. PreparationRefer to figure 3-4. The relay contact terminals and power supply connections a
8 IM 12D08N05-01EHigh voltage sectionContact(S1, S2)outputcablesmAcablesContact(S3, S4)outputcablesSensorCablesInputcontactPowercableSuitable for cabl
9IM 12D08N05-01E3 INSTALLATION AND WIRING3-3. Wiring the power supply3-3-1. General precautionsMake sure the power supply is switched off. Also, make
10 IM 12D08N05-01E3-3-5. Grounding the housingForthesafetyoftheuserandtoprotecttheinstrument against interference, the housing must always b
11IM 12D08N05-01E3 INSTALLATIONANDWIRING3-4. Wiring the contact signals3-4-1. General precautionsThe contact output signals consist of volt-age-fre
EnterMeasurement setup Temperature settings Temp. Compensation Calibration settings Concentration Measure Conductivity only Configure sens
12 IM 12D08N05-01EFigure 3-9. Sensor wiring diagrams3-6. Wiring of sensorsGeneral precautionsGenerally,signalsfromsensorsareatlowvolt-age and
13IM 12D08N05-01E3 INSTALLATIONANDWIRING3-6-1. Sensor cable connections using junc-tion box (BA10) and extension cable (WF10)Where a convenient in
14 IM 12D08N05-01EExtension cable may be purchased in bulk quantities or in pre-finished lengths. In the case of bulk quantities cut to length, then i
15IM 12D08N05-01E4 OPERATION OF EXAxt SC450G4-1. Main display functionsGo to trend screenGo to zoom screenGo to status screenGo to maintenance screen2
16 IM 12D08N05-01E4-3-1. Actual mA1=thecurrentoutputinmAofthe first current output, which is defined as mA1. The range and function of this mA
17IM 12D08N05-01E4OPERATIONOFEXAxtSC450G4-3-15. HART Device revisionSometimesthefirmwareofadeviceisupdated in a way that the communication
18 IM 12D08N05-01E“RETURNKEY”exittopreviousdisplayMain displayPrimary setup displayCommisioning menu displayInstrumentinHOLD4-7. Navigation of
19IM 12D08N05-01E4OPERATIONOFEXAxtSC450G
20 IM 12D08N05-01EMeasurement setupMain parameterChoose the required parameter, either con-ductivity or resistivity. If the main parameter is changed
21IM 12D08N05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.ConfigureSensor Cellconstant 0.1cm-1 0.005
Electrostatic dischargeThe EXAxt converter contains devices that can be damaged by electrostatic discharge. When servicing this equipment, please obse
22 IM 12D08N05-01E5-4. Calibration settingsAir adjust limitTo avoid cable influences on the measurement, a “zero” calibration with a dry sensor may be
23IM 12D08N05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.Calibration Airadjust 10.00µS 0µS 20µS
24 IM 12D08N05-01E5-6. mA output setupThe general procedure is to first define the function (control, output, simulate, off) of the output and second
25IM 12D08N05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.mA1 (control) Expire time 0.0 sec. 0 sec.
26 IM 12D08N05-01EExpire timeIftheoutputisover100%forlongerthantheexpiretime,theoutputwillreturnto0%.Damping timeThe response to a st
27IM 12D08N05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.PID-controlS1 Setpoint 250.0µS/cm -inf
28 IM 12D08N05-01E5-8. FailA fail contact is energized when a fail situation occurs.Failsituationsareconfiguredinsecton5-12.ForSOFTFailsthe
29IM 12D08N05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.Simulation Percentage 50% 0% 100%S2, S3, S
30 IM 12D08N05-01E5-12. Error configurationErrors 1/3 ~ 3/3Errors are intended to notify the user of any unwanted situations. The user can determine w
31IM 12D08N05-01E5MENU STRUCTURECOMMISSIONINGEnterErrors 2/3Temperature too highTemperature too lowPolarization detectCalibr. time exceededUSP li
ii The following safety symbols are used on the product as well as in this manual.DANGERThis symbol indicates that an operator must follow the instru
32 IM 12D08N05-01E5-14. Advanced setupDefaultsThe functionality of the EXAxt allows to save and load defaults to come to a known instru-ment setting.
33IM 12D08N05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values Low HighHART Network address 0 0 15
34 IM 12D08N05-01E5-15. Display setupMain displayThe main display consists of three lines with Process Values. Each line is user definable with the re
35IM 12D08N05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values Low HighY-axis Conduct low 0 µS/cm -inf +infY-ax
36 IM 12D08N05-01E6-1. GeneralThe nominal cell constant of a conductiv-ity sensor is determined at the construction stage, because it is a factor set
37IM 12D08N05-01E6CALIBRATION6-7. Temperature calibration In order to make the most accurate measure-ments, it is important to have a precise temper-
38 IM 12D08N05-01E7-1. Periodic maintenance The converter requires very little periodic main-tenance, except to make sure the front window is kept cle
39IM 12D08N05-01E7 MAINTENANCETouchscreen adjustment A few years after using, the touchscreen may deviate from the correct position due to aging de
40 IM 12D08N05-01E8-1. General The EXAxt is a microprocessor-based analyzer that performs continuous self-diagnostics to verify that it is working cor
41IM 12D08N05-01E9QUALITYINSPECTION9. QUALITY INSPECTIONQualityInspectionStandards SC450GConductivity or Resistivity Converter QIS 12D08N05-01E1s
IM 12D08N05-01EiiiIM 12D08N05-01E4th Edition: Mar. 2012 (YK) All Rights Reserved, Copyright © 2007, Yokogawa Electric CorporationTABLE OF CONTENTSPREF
42 IM 12D08N05-01E2/3 QIS 12D08N05-01EThis test is done on the “HIF” display of “Factory Mode”. a. Touch the [Setup] icon. b. Touch the [Commissioni
43IM 12D08N05-01E9QUALITYINSPECTION3/3 QIS 12D08N05-01EAfter all tests are completed, a. Touch the [Exit] twice to return to the “Service” display
44 IM 12D08N05-01E
45IM 12D08N05-01E9SPAREPARTS10. SPARE PARTSSeeCustomerMaintenancePartsList.
46 IM 12D08N05-01EAppendix 1Temperature compensationThe conductivity of a solution is very dependent on temperature. Typically for every 1°C change in
47IM 12D08N05-01EAAPPENDICESB. Calculation of temperature coefficient factor(with two known conductivity values at different temperatures)Measure the
48 IM 12D08N05-01ECheckingWhen the temperature coefficient already set is accurate, the conductivity to be displayed must be constant regardless of li
49IM 12D08N05-01EAAPPENDICESThe matrix can be cleared before entering new values. Next new matrix values can be entered as described above. The EXAxt
50 IM 12D08N05-01EAppendix 2TDS readingsTheconceptofTotalDissolvedSolids(TDS)iswidely used as an indication of the total solute concentration
51IM 12D08N05-01EAAPPENDICESAppendix 3Calibration solutions for conductivityNote! This section should be read in conjunction with the calibration sect
IM 12D08N05-01Eiv 4. OPERATION OF EXAxt SC450G ...154-1. Main display funct
52 IM 12D08N05-01EIfitismoreconvenient,theusermaymakesolutionsfromSodiumChloride(NaClorcommontable salt) with the help of the followin
53IM 12D08N05-01EAAPPENDICESAppendix 4Sensor selection for the EXAxt SC450GYokogawa supplies the following sensors, and their application is briefly d
54 IM 12D08N05-01ESC42-EP04 cc = 10 cm-1 2-electrode sensorThis sensor is in the program for historical reasons. It makes little sense to select
55IM 12D08N05-01EAAPPENDICESAppendix 5, How to enter the cell constant for the SC21¨G, SC8SG, and SC4AJWhen install a new sensor, or replace a new one
56 IM 12D08N05-01EOnline menu Level 1 menu Level 2 menu Level 3 menu Level 4 menuProcess valuesPrimary value (SC/RES/Conc.)Secondaryvalue(Temp.)Tert
57IM 12D08N05-01EAAPPENDICESOnline menu Level 1 menu Level 2 menu Level 3 menuLevel 4 menuMost appl. ErrorCalibrationHold InstrumentError description
58 IM 12D08N05-01EOnline menu Level 1 menu Level 2 menu Level 3 menu Level 4 menuCommissioningOutput setup mA1 setupmA2 setup (similar to mA1) Type =
59IM 12D08N05-01EAAPPENDICESOnline menu Level 1 menu Level 2 menu Level 3 menu Level 4 menuLoop testBasic setup TagDistributorModelDevice information
60 IM 12D08N05-01EAppendix 7, Control drawing for FM approval
CMPL12D08N05-02ECustomerMaintenanceParts List©Copyright 2007, 4thEdition:Oct,2011(YK)Subjecttochangewithoutnotice.Model SC450G [Style: S2]Con
IM 12D08N05-01Ev6. CALIBRATION ...366-1.Gener
CMPL12D08N05-02E24thEdition:Oct.2011(YK)2345311078963Pipe/Wall Mounting Hardware (Option code: /U)Awning Hood (Option code: /H5)Panel Mounting H
Revision RecordManual Title : Model SC450G [Style: S2] Conductivity/Resistivity Converter Manual Number : IM 12D08N05-01E Edition Date Remark (s)1
1IM 12D08N05-01E1 INTRODUCTION AND GENERAL DESCRIPTIONThe Yokogawa EXAxt SC450G is a converter designed for industrial process monitoring, mea-sureme