Wiring the Pulse Input Module and Digital Input Modules (-D05, -D24)
Note
With the pulse input module and digital input module, the (–) terminal and unassigned terminals on all channels are
shorted internally.
When the screw terminal plate (model 772080) is connected to the pulse input module and digital input module, the
terminal arrangement differs from that of clamp terminals, therefore wire according to the markings on the terminal
cover. Do not connect anything to the b terminals.
Wiring the Pulse Input and Digital Input (-D05) Modules
• Transistor input
• Contact input
• 5-V logic input
Main Input Specifications (for Pulse Input and Digital Input (-D05))
Input type: DI (non-voltage contact, open collector, and 5 V logic)
Input format: Pull-up at approximately 5 V/approximately 5 k, common electric
potential between channels
Minimum detection pulse width:
Pulse input: 4 µs
Digital input (-D05): Twice the measurement interval or more
Input threshold level: Pulse input
Non-voltage contact or open collector:
Counts when changing from contact open (100 k or more) to
contact close (100 or less)
5-V logic: Counts when changing from 1 V or less to 3 V or more
Digital input (-D05)
Non-voltage contact or open collector:
ON at 100 or less and OFF at 100 k or more
5-V logic: OFF at 1 V or less and ON at 3 V or more
Contact/transistor rating:
Contact with a rating of 15 VDC or greater and 30 mA or greater
Vce and Ic are transistors with ratings of 15 VDC or more, and 30
mA or more, respectively.
Terminal type: Clamp
Applicable wire size: 0.14 to 1.5 mm
2
(AWG26 to 16)
Wiring the Digital Input (-D24) Module
24 V
• 24-V logic input
Main Input Specifications (for Digital Input (-D24))
Input type: DI (24-V logic)
Input format: Common potential between channels
Minimum detection pulse width:
Twice the sampling interval or more
Input threshold level: 24-V logic: OFF at 6 V or less and ON at 16 V or greater
Terminal type: Clamp
Applicable wire size: 0.14 to 1.5 mm
2
(AWG26 to 16)