Yokogawa PC-Based MX100 Bedienungsanleitung Seite 63

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 76
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 62
5-3
IM MW180-01E
1
2
3
4
5
App
Index
Troubleshooting
The connected input/output module is not detected.
Probable Cause Corrective Action/Description Reference Page
Module connection or module startup error.
Attached the module while the power was ON.
Turn OFF the power. Detach the input/output module
once and attach it again.
*
The PC software, main module, and input/output
module style and release number rule is not
upheld.
Check the PC software release number and main
module style number before upgrading the version.
[PC software release no.] ≥ [main module style no.] ≥
[input/output module style no.]
**
Carried out an incorrect calibration. Recalibrate. 4-2
* See the MW100 Data Acquisition Unit Users Manual (IM MW100-01E).
** See the MW100 Data Acquisition Unit Operation Guide (IM MW100-01E).
5.1 Troubleshooting