IM 12D06D05-01E4th EditionUser’sManualModel ISC450G [Style: S2]Inductive Conductivity Converter

Seite 2 - Measurement setup

2 IM 12D06D05-01E2. GENERAL SPECIFICATIONS OF EXAxt ISC450GA) Input specifications :CompatiblewiththeYokogawainductiveconductivityISC40serie

Seite 3

3IM 12D06D05-01E2 GENERALSPECIFICATIONS Hold : ContactcanbeusedtosignaltheHOLDsituation. Failsafe : ContactS4isprogrammedasfail-sa

Seite 4

4IM 12D06D05-01EN) Safety and EMC conforming standards , Safety : EN61010-1   CSAC22.2No.61010-1   UL61010-1   FM3611ClassI,

Seite 5 - 4th Edition: Mar. 2012 (YK)

5IM 12D06D05-01E3 INSTALLATIONANDWIRING3-1. Installation and dimensions3-1-1. Installation siteTheEXAxt450converterisweatherproofandcan be i

Seite 6

6 IM 12D06D05-01Emin.185 (7.25")min.195(7.75")138(5.43")138(5.43")M6M6M5138(5.43")+1 0138(5.43")+1 0Figure 3-2. Opt

Seite 7

7IM 12D06D05-01E3 INSTALLATIONANDWIRING3-2. Wiring3-2-1. PreparationRefertofigure3-4.Therelaycontactterminalsand power supply connections a

Seite 8

8 IM 12D06D05-01EHigh voltage sectionContact(S1, S2)outputcablesmAcablesContact(S3, S4)outputcablesSensorCablesInputcontactPowercableSuitable for cabl

Seite 9 - 4-20mA(mA1: +HART)

9IM 12D06D05-01E3 INSTALLATIONANDWIRING3-3. Wiring the power supply3-3-1. General precautionsMake sure the power supply is switched off. Also, make

Seite 10

10 IM 12D06D05-01E3-3-5. Grounding the housingForthesafetyoftheuserandtoprotecttheinstrument against interference, the housing must always b

Seite 11 - GENERALSPECIFICATIONS

11IM 12D06D05-01E3 INSTALLATIONANDWIRING3-4. Wiring the contact signals3-4-1. General precautionsThe contact output signals consist of volt-age-fre

Seite 12

EnterMeasurement setup Temperature settings Temp. Compensation Calibration settings Concentration Measure Conductivity only Configure sens

Seite 13 - 3. INSTALLATION AND WIRING

12 IM 12D06D05-01EFigure 3-9a. DimensionsIt is necessary to use screening/shielding on the output signal cables. Terminal 63 is used to connect the s

Seite 14

13IM 12D06D05-01E3 INSTALLATIONANDWIRINGFigure 3-9b. Dimensions3-6-4. Sensor cable connections using junc-tion box (BA10) and extension cable (WF1

Seite 15

14IM 12D06D05-01EExtension cable may be purchased in bulk quantities or in pre-finished lengths. In the case of bulk quantities cut to length, then i

Seite 16 - High voltage section

15IM 12D06D05-01E3 INSTALLATIONANDWIRINGFigure 3-11.a. Figure 3-11.b. Figure 3-11.c.

Seite 17 - Unit: mm(inch)

16 IM 12D06D05-01E4-1. Main display functionsGo to trend screen Go to zoom screenGo to status screenGo to maintenance screenTag :EXAxt ISC4503.76025.

Seite 18 - FUSE: 2A/250 VAC/T

17IM 12D06D05-01E4OPERATIONOFEXAxtISC450G4-3-1. Actual mA1 = the current output in mA of the first current output, which is defined as mA1. The ra

Seite 19

18 IM 12D06D05-01E4-3-14. HART Device revisionSometimesthefirmwareofadeviceisupdated in a way that the communication file (HART DD) need revisi

Seite 20

19IM 12D06D05-01E4OPERATIONOFEXAxtISC450G“RETURNKEY”exittopreviousdisplayMain displayPrimary setup displayCommisioning menu displayInstrument

Seite 21 - UNIT: mm (inch)

20 IM 12D06D05-01E5-1. Configure sensor Measuring unit /cm /mEither /cm or /m can be chosen here. The Proc-ess values will be expressed in S/cm or S/m

Seite 22 - / CONVERTER

21IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.ConfigureSensor Cellconstant 1.88cm-1 0.2

Seite 23 - Figure 3-11.c

Electrostatic dischargeThe EXAxt converter contains devices that can be damaged by electrostatic discharge. When servicing this equipment, please obse

Seite 24 - Tag :EXAxt ISC450

22 IM 12D06D05-01E5-4. Calibration settingsAir adjust limitTo avoid cable influences on the measurement, a“zero”calibrationwithadrysensormaybe

Seite 25 - OPERATIONOFEXAxtISC450G

23IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.Calibration Airadjust 100.00µS 0µS  100

Seite 26

24IM 12D06D05-01E5-6. mA output setupThe general procedure is to first define the function (control, output, simulate, off) of the output and second

Seite 27 - Commisioning menu display

25IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.mA1 (output) Damping time 0.0 sec. 0.0 sec.

Seite 28

26 IM 12D06D05-01EExpire timeIftheoutputisover100%forlongerthantheexpiretime,theoutputwillreturnto0%.Damping timeThe response to a st

Seite 29

27IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.AlarmS1 Setpoint 900.0mS/cm(high) -inf 

Seite 30

28 IM 12D06D05-01E5-8. FailA fail contact is energized when a fail situation occurs.Failsituationsareconfiguredinsecton5-11.ForSOFTFailsthe

Seite 31

29IM 12D06D05-01E5MENU STRUCTURECOMMISSIONING Menu Parameter Default Range values min. max.Simulation Percentage 50% 0%  100%S2,S3,

Seite 32

30 IM 12D06D05-01E5-11. Error configurationErrors 1/3 ~ 3/3Errors are intended to notify the user of any unwanted situations. The user can determine w

Seite 33

31IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.Errors Highlimit 1.000S 0.1S  3SErrors Lo

Seite 34

ii The following safety symbols are used on the product as well as in this manual.DANGERThis symbol indicates that an operator must follow the instruc

Seite 35

32 IM 12D06D05-01E5-13. Advanced setupDefaultsThe functionality of the EXAxt allows to save and load defaults to come to a known instru-ment setting.

Seite 36

33IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values Low HighHART Network address 0 0 15

Seite 37

34IM 12D06D05-01E5-14. Display setupMain displayThe main display consists of three lines with Process Values. Each line is user definable with the re

Seite 38

35IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values Low HighY-axis Conductlow 0µS/cm -inf  +infY-ax

Seite 39

36 IM 12D06D05-01E6. CALIBRATION6-1. Before calibration1. When is calibration necessary?Calibration of con duc tiv i ty in stru ments is not nor

Seite 40

37IM 12D06D05-01E6CALIBRATIONCell constant: The nominal cell constant of the sensor is 1.88 cm-1forthePEEKsensortypes and 3.00 cm-1forthePFA

Seite 41

38 IM 12D06D05-01E6-2. Cell constant manualThe intention of this calibration routine is to fine tune a sensor for which only the nominal cell constant

Seite 42

39IM 12D06D05-01E6CALIBRATION6-7. Temperature calibration In order to make the most accurate measure-ments, it is important to have a precise temper-

Seite 43

40IM 12D06D05-01E7-1. Periodic maintenance The converter requires very little periodic main-tenance, except to make sure the front window is kept cle

Seite 44 - 6. CALIBRATION

41IM 12D06D05-01E7 MAINTENANCETouchscreen adjustment A few years after using, the touchscreen may deviate from the correct position due to aging de

Seite 45 - Only for PEEK

IM 12D06D05-01EiiiIM 12D06D05-01E4th Edition: Mar. 2012 (YK) All Rights Reserved, Copyright © 2007, Yokogawa Electric CorporationTABLE OF CONTENTSPREF

Seite 46

42IM 12D06D05-01E8-1. General The EXAxt is a microprocessor-based analyzer that performs continuous self-diagnostics to verify that it is working cor

Seite 47 - CALIBRATION

43IM 12D06D05-01E9QUALITYINSPECTION9. QUALITY INSPECTIONQualityInspectionStandards ISC450GInductive Conductivity Converter QIS 12D06D05-01E1st Ed

Seite 48 - 7. MAINTENANCE

44IM 12D06D05-01E2/3 QIS 12D06D05-01EThis test is done on the “HIF” display of “Factory Mode”. a. Touch the [Setup] icon. b. Touch the [Commissioni

Seite 49 - 7 MAINTENANCE

45IM 12D06D05-01E9QUALITYINSPECTION3/3 QIS 12D06D05-01ETable 4 Data Display Current Output (mA DC) 4 4 ±0.02 12 12 ±0.02 20 20 ±0.02 After all t

Seite 50 - 8. TROUBLESHOOTING

46IM 12D06D05-01E

Seite 51 - Standards

47IM 12D06D05-01E10SPAREPARTS10. SPARE PARTSSeeCustomerMaintenancePartsList.

Seite 52

48IM 12D06D05-01EAPPENDIX 1Temperature compensationTheconductivityofasolutionisverydependentontemperature.Typicallyforevery1°Cchangein

Seite 53

49IM 12D06D05-01EAAPPENDICESB. Calculation of temperature coefficient factor(with two known conductivity values at different temperatures)Measure the

Seite 54

50 IM 12D06D05-01E• CheckingWhen the temperature coefficient already set is accurate, the conductivity to be displayed must be constant regardless of

Seite 55 - 10. SPARE PARTS

51IM 12D06D05-01EAAPPENDICESThe matrix can be cleared before entering new values. Next new matrix values can be entered as described above. The EXAxt

Seite 56 - APPENDICES

IM 12D06D05-01Eiv 4. OPERATION OF EXAxt ISC450G ...164-1.Maindisplayfunct

Seite 57

52 IM 12D06D05-01EAPPENDIX 3Calibration solutions for conductivityNote! This section should be read in conjunction with the calibration section (Chapt

Seite 58

53IM 12D06D05-01EAAPPENDICESIf it is more convenient, the user may make solutionsfromSodiumChloride(NaClorcom-mon table salt) with the help of t

Seite 59

54IM 12D06D05-01EOnline menu Level 1 menu Level 2 menu Level 3 menu Level 4 menuProcess valuesPrimary value (SC/Conc.)Secondaryvalue(Temp.)Tertiary

Seite 60

55IM 12D06D05-01EAAPPENDICESOnline menu Level 1 menu Level 2 menu Level 3 menuLevel 4 menuMost appl. ErrorCalibrationError description / remedyCCCali

Seite 61

56 IM 12D06D05-01EOnline menu Level 1 menu Level 2 menu Level 3 menu Level 4 menuCommissioningOutput setup mA1 setupmA2 setup (similar to mA1)Type = c

Seite 62

57IM 12D06D05-01EAAPPENDICESOnline menu Level 1 menu Level 2 menu Level 3 menu Level 4 menuLoop testBasic setup TagDistributorModelDevice information

Seite 63

58 IM 12D06D05-01EAPPENDIX 6, Control drawing for FM approval

Seite 64

CMPL12D06D05-02ECustomerMaintenanceParts List©Copyright 2007, 4thEdition:Oct.,2011(YK)Subjecttochangewithoutnotice.Model ISC450G [Style: S2]I

Seite 65

CMPL12D06D05-02E24thEdition:Oct..2011(YK)2345311078963Pipe/Wall Mounting Hardware (Option code: /U)Awning Hood (Option code: /H5)Panel Mounting

Seite 66

Revision RecordManualTitle: ModelISC450G[Style:S2]InductiveConductivityConverter Manual Number : IM 12D06D05-01E Edition Date Remark (s)1s

Seite 67 - Parts List

IM 12D06D05-01Ev6. CALIBRATION ...366-1.Befor

Seite 69 - Revision Record

1IM 12D06D05-01E1 INTRODUCTIONANDGENERALDESCRIPTION1. INTRODUCTION AND GENERAL DESCRIPTIONTheYokogawaEXAxtISC450Gisaconverterdesigned for i