Yokogawa FlowNavigator Software Bedienungsanleitung Seite 6

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 7
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 5
<A1. Header>
6
IM 01C25R51-10E
Appendix A Flow Parameters Download Procedure
The basic procedure for downloading the ow parameters to a device is as follows. Refer to Figure A-1, A-2.
Step1
Upload from Device
Perform “Upload from Device” in the Ofine Parameter screen so that the setting parameters (“Device Flow Parameters” and
“Device General Parameters”) in the device are saved to the ofine database on your PC.
The parameters will then be displayed in the Ofine Parameter screen.
Step2
Flow Conguration
Perform Flow conguration in accordance with the guidance in the Flow Conguration Wizard. “Device Flow Parameters” are
output by Flow Conguration Wizard.
Save “Device Flow Parameters” as XMV le (EJXMVTool) or VMV le (DYFMVTool) in the last menu, and click the “Apply
button.
“Device Flow Parameters” are then stored to the ofine database on your PC.
NOTE
Clicking on Apply does not download the ow parameters to the device.
Step3
Download to Device
Perform “Download to Device” in the Ofine Parameter screen. “Device Flow Parameters” will then be downloaded together
with the “Device General Parameters” to the device.
IMPORTANT
For FOUNDATION Fieldbus, following block must be checked on the “Download blocks” menu in the Ofine Parameter screen
to download “Device Flow Parameters” set by Flow Conguration Wizard.
EJX910 FF DTM: FTB block, DYF(SoftDL) FF DTM: AR block
“Download blocks” menu (EJX910 FF DTM)
“Download blocks” menu (DYF(SoftDL) FF DTM)
Figure A-1 “Download blocks” menu in the Ofine Parameter
Seitenansicht 5
1 2 3 4 5 6 7

Kommentare zu diesen Handbüchern

Keine Kommentare