ote 4. Special Condition for Safe Use
WARNING
Electrostatic charge may cause an explosion
hazard. Avoid any action that cause the gen-
eration of electrostatic charge, such as rubbing
with a dry cloth on coating face of the product.
Associated
Apparatus
(Zone 2 only)
Transmitter
Supply
Nonhazardous Location
[Installation Diagram]
Hazardous Location
+ 
+ 
F0209.EPS
Ratings of the Power Supply as follows;
Maximum Voltage: 30 V
ATEX Type of Protection “Dust”
[When combined with II 2G]
Follow the installation instructions described in 'b.
ATEX Flameproof Type.'
• Applicable Standard: EN61241-0:2006, EN61241-
1:2004
• Type of Protection and Marking Code: II 2D Ex tD
A21 IP6X
• Maximum Surface Temperature:
T85°C (Tamb.: –40* to 75°C, Tp:85°C), T100°C
(Tamb.: –40* to 80°C, Tp:100°C), and T120°C
(Tamb.: –40* to 75°C, Tp:120°C)
*–15°C when /HE is specified.
Note 1. Installation instructions
The cable entry devices and blanking elements
shall be of a certificated type providing a level
of ingress protection of at least IP6X, suitable
for the conditions of use and correctly installed.
[When combined with II 1G]
Follow the installation instructions described in 'a.
CENELEC ATEX Intrinsically Safe Type.'
• Applicable Standard: EN50281-1-1:1997
• Type of Protection and Marking Code: II 1D
• Maximum Surface Temperature:
T65°C (Tamb.: 40°C), T85°C (Tamb.: 60°C), and
T105°C (Tamb.: 80°C)
Note 1. Installation instructions
The cable entry devices and blanking elements
shall be of a certificated type providing a level
of ingress protection of at least IP6X, suitable
for the conditions of use and correctly installed.
WARNING
OPEN CIRCUIT BEFORE REMOVING
COVER. INSTALL IN ACCORDANCE WITH
THIS USER’S MANUAL
Take care not to generate mechanical sparking
when access to the instrument and peripheral
devices in hazardous locations.