Operation
Starting Memory Sampling
1.
Press START once.
On a DX with advanced security (/AS1 option; release number 4 or later) display
the start recording screen
*
. With START selected, press DISP/ENTER once.
Memory sampling starts.
* This is because the Batch function is enabled upon shipment from the factory.
START key
Memory sampling progress
DISP: Display data
EVENT: Event data
Recording data
Operation complete.
Stopping Memory Sampling
1.
Press STOP once.
Display the confirmation
window.
2.
Select Mem+Math or Memory using the left and right arrow keys.
Memory: Stops memory sampling.
Mem+Math: Stops memory sampling and computation (option).
On models without the computation function (option), the confirmation message
“Do you want to stop data storage?” appears. Select Yes
.
3.
Press DISP/ENTER once.
Stop memory sampling.
Operation complete.