Yokogawa ADMAG AXF Bedienungsanleitung Seite 96

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 133
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 95
IM 01E20D01-01E
6-34
6. OUTLINE
Table of Optional Specifications (Size 500 mm (20 in.) to 2600 mm (104 in.)) (continued)
AXF***W-N
Submersible
AXF***G-N
General
Applicable Model
Item
Specifications
Five-point Calibration
in User-specified
Span
A flow test near 0, 25, 50, 75, and 100% of the user-specified span is performed instead of the flow test of
the standard 2m/s span and a test certificate (QIC) is submitted. Specify the span (100% flow span) whose
corresponding flow velocity lies between 0.5 to 10 m/s and that is less than the maximum line capacity.
Selectable range of flow rate span is showing below.
SC
T26-6.EPS
Code
Remote
Flowtube
Remote
Flowtube
: Available –: Not available
Selectable range of flow rate span : m
3
/h
(Flow rate span velocity : m/s)
354 (0.5) to 7068 (10.00)
509 (0.5) to 8200 (8.06)
693 (0.5) to 8200 (5.92)
905 (0.5) to 8200 (4.53)
1146 (0.5) to 8200 (3.58)
1414 (0.5) to 8200 (2.90)
1711 (0.5) to 8200 (2.40)
2036 (0.5) to 8200 (2.01)
2577 (0.5) to 30000 (5.82)
3181 (0.5) to 30000 (4.71)
3620 (0.5) to 30000 (4.14)
4581 (0.5) to 30000 (3.27)
5655 (0.5) to 30000 (2.65)
6843 (0.5) to 30000 (2.19)
8143 (0.5) to 30000 (1.84)
9557 (0.5) to 30000 (1.57)
Size : mm
(in.)
500 (20)
600 (24)
700 (28)
800 (32)
900 (36)
1000 (40)
1100 (44)
1200 (48)
1350 (54)
1500 (60)
1600 (64)
1800 (72)
2000 (80)
2200 (88)
2400 (96)
2600 (104)
Hydrostatic Test
Level 2: The Declaration and the Calibration Equipment List are issued.
L2
Calibration Certificate
Level 3: The Declaration and the Primary Standard List are issued.
L3
Level 4: The Declaration and the Yokogawa Measuring Instruments Control System are issued.
L4
The test verifies the absence of leaks by applying the following water pressures (which are determined
under process connection conditions) to lining for ten minutes. Test results are described in the Note column
of a test certificate(QIC).
Process Connection: Water Pressure:
JIS 10K, ANSI Class 150, DIN PN10 1.5 MPa
JIS F12 1.25 MPa
T01