User’sManualRModel OX400Low Concentration Zirconia Oxygen Analyzer[Style: S2]IM 11M10B01-01EIM 11M10B01-01E3rd Edition
<1.Outline>1-1IM 11M10B01-01E1. OutlineTheOX400isahighlyaccurateandreliablelowconcentrationzirconiaoxygenanalyzerthatiscapa
<1.Outline>1-2IM 11M10B01-01E
<2.Specications>2-1IM 11M10B01-01E2. Specications2.1 StandardSpecicationsMeasurement object:Oxygenconcentrationsininertgasescon
<2.Specications>2-2IM 11M10B01-01EContact output: 3 outputsError contact, O2concentrationhigh/lowalarmcontact,rangemarkeroutput Mu
<2.Specications>2-3IM 11M10B01-01ELine connectionGasinlet:Rc1/4or1/4NPTfemaleGasoutlet:Rc1/4or1/4NPTfemaleElectricalconnect
<2.Specications>2-4IM 11M10B01-01EResponse time : 90% response: Within 10 sec (0-1% or more): Within 30 sec (less than 0-1%)Drift:±2%FS/we
<2.Specications>2-5IM 11M10B01-01EConsumablesItem Partno.andrating Qt’yActivatedlterelementkit(15timesreplacement) K9643KK 1Filt
<2.Specications>2-6IM 11M10B01-01EPanelmounttypewithaspirator(OX400--A--/P)40 to 50(Line connection : in case of Rc1/4)(Line connecti
<2.Specications>2-7IM 11M10B01-01EDesktoptypewithbuilt-inpumpornosuctiondevice(OX400----)10 to 2021313220350 19.3RearNote: P
<2.Specications>2-8IM 11M10B01-01EDesktoptypewithaspirator(OX400--A--)With multi selector functionRear40 to 5060 to 70(Line connection
iIM 11M10B01-01E<INTRODUCTION>INTRODUCTIONn Safety,Protection,andModicationoftheProduct• In order to protect the system controlled by t
<2.Specications>2-9IM 11M10B01-01EActivatedcarbonlter(Rc1/4 or 1/4NPT)(Rc1/4 or 1/4NPT)110K9643KH: Rc1/4K9643KJ: 1/4NPT± 0.5± 0.5Unit:
<2.Specications>2-10IM 11M10B01-01E2.6 PipingDiagramPPFlow meterSuction pumpHeaterOxygensensorINOUTINOUTOxygensensorWith aspirator (Sampling
<2.Specications>2-11IM 11M10B01-01E2.7 WiringDiagramMS1MS2MS3MSI1MSI2MSICOMMSCOM1st flow path2nd flow path3rd flow pathOX400 Rear Termin
<2.Specications>2-12IM 11M10B01-01E
<3.Installation>3-1IM 11M10B01-01E3. Installation3.1 InstallationLocationThe OX400 is a measuring instrument intended to be installed in
<3.Installation>3-2IM 11M10B01-01E3.2 HowtoInstall3.2.1 InstallingDesktopTypePlaceandoperatetheOX400onalevelsurfaceasshowninF
<3.Installation>3-3IM 11M10B01-01E3.2.2 InstallingPanelMountType(1) Attach a panel mount frame to the side of the OX400, insert it in
<3.Installation>3-4IM 11M10B01-01E3.2.3 MountingActivatedCarbonFilterMountthelteronapanelorwall.ToxthelterwithM5screws.Wh
<4.PipingandWiring>4-1IM 11M10B01-01E4. PipingandWiring4.1 PipingBe sure to observe the following precautions when connecting the gas
<4.PipingandWiring>4-2IM 11M10B01-01E!INLETOUTLETN200C USR172608 SERIAL(RS-232)FUSE250V T3.15A!RCCOMRC3RC2RC1FAILDOPUMP OFFGV-+mA-+MS1MS2MS3M
iiIM 11M10B01-01E<INTRODUCTION>n Notes on Hardwarel Appearance and AccessoriesCheck the following when you receive the product:• Appearance•
<4.PipingandWiring>4-3IM 11M10B01-01EMS1MS2MS3MSI1MSI2MSICOMMSCOM1st flow path2nd flow path3rd flow pathOX400 Rear Terminal (M3 screw)The
<4. Piping and Wiring>4-4IM 11M10B01-01E1) Primary output terminal This is a terminal to output oxygen concentration at a 4-20 mA DC current.
<4.PipingandWiring>4-5IM 11M10B01-01ETable4.3 MeasurementFlowPathSwitchingOutputbyMulti-selectorMS1-MSCOM MS2-MSCOM MS3-MSCOMMS O
<4.PipingandWiring>4-6IM 11M10B01-01E
<5.NamesandFunctions>5-1IM 11M10B01-01E5. NamesandFunctions5.1 FrontPanel123456891011127Figure5.1DescriptionofFrontPanel1Main
<5.NamesandFunctions>5-2IM 11M10B01-01E8Arrow Keys [◄],[►]:Pressthekeystomovethroughthedigitsofthenumbertobeset.[▲],[▼]:Pr
<5.NamesandFunctions>5-3IM 11M10B01-01E5.2 RearPanel!INLETOUTLETN200C USR172608 SERIAL(RS-232)FUSE250V T3.15A!RCCOMRC3RC2RC1FAILDOPUMP
<5.NamesandFunctions>5-4IM 11M10B01-01E11ExternalOutputTerminal(SecondaryOutput) Measured values are output at the set voltage (0-1, 0-
<6.Operation and Parameters>6-1IM 11M10B01-01E6. OperationandParametersThe OX400 has two modes, “Measurement Mode” and “Maintenance Mode
<6.OperationandParameters>6-2IM 11M10B01-01E6.1 StartupandSettingsTurningonthePowerTurn on the POWER switch on the front panel. After a
iiiIM 11M10B01-01E<INTRODUCTION>• Display contents on the display Representedby“”. Example:Maindisplay--> “HEAt” Example:
<6.Operation and Parameters>6-3IM 11M10B01-01E%%%%%%%%%10 0.0Fr.HIPPMPressing the [MODE] key in the event of an input error returns to prev
<6.OperationandParameters>6-4IM 11M10B01-01E6.1.2 SettingSecondaryOutput“oUt2”Operation:Holddownthe[MODE]keyfor2seconds,select“
<6.Operation and Parameters>6-5IM 11M10B01-01E6.1.4 SettingBurnoutFunction“nAMU”Operation:Holddownthe[MODE]keyfor2seconds,sele
<6.OperationandParameters>6-6IM 11M10B01-01E6.1.5 SettingAlarms(OxygenConcentrationHigh/LowAlarms)“ALM”Operation:Holddownthe[MODE]
<6.Operation and Parameters>6-7IM 11M10B01-01E6.1.6 SettingCalibrationGasConcentration“SEt.C”Operation:Holddownthe[MODE]keyfor2
<6.OperationandParameters>6-8IM 11M10B01-01E6.1.8 SettingOutputSmoothing“SMoo”Iftheoxygenconcentrationofsamplegasrapidlychanges,
<6.Operation and Parameters>6-9IM 11M10B01-01Eandpressthe[ENT]key.Checkthecurrentcalibrationcoefcient.Thesecoefcientsareupda
<6.OperationandParameters>6-10IM 11M10B01-01ErEV▼↓↑▲MODECAL▼↓↑▲▼↓↑▲[ENT]1. 8rEVPPM[ENT]▼↓↑▲%To “rEV”Hold down for 2 seconds.Return to “rEV.”F
<6.Operation and Parameters>6-11IM 11M10B01-01Ecalibration. However, there are the following limits to the O2 concentration. Be sure to per
<6.OperationandParameters>6-12IM 11M10B01-01E6.3 CommunicationTheOX400hasRS232serialcommunicationasstandard.Oxygenconcentration,al
ivIM 11M10B01-01E<INTRODUCTION>• When opening the cover to replace the sensor, turn off the power switch, remove the power plug from the socket
<6.Operation and Parameters>6-13IM 11M10B01-01ETable6.3 ErrorFactorandOutputCharacterListOutputcharacters Errorfactor0001 Err1: Se
<6.OperationandParameters>6-14IM 11M10B01-01E6.4 MeasurementGasSamplingUsingAspiratorIf you selected “-A”: with an aspirator as the samp
<6.Operation and Parameters>6-15IM 11M10B01-01E6.5 SwitchingMeasurementFlowPathUsingMulti-SelectorThisitemisanoption,soifyoud
<6.OperationandParameters>6-16IM 11M10B01-01E6.6 NumericalValue,DecimalPoint,andUnitInputOperationFigure6.21showsanexampleofope
<7.InspectionandMaintenance>7-1IM 11M10B01-01E7. InspectionandMaintenanceRoutine inspection and maintenance is important to ensure ope
<7.InspectionandMaintenance>7-2IM 11M10B01-01E7.2.1 InspectionintheEventofanAlarm(ALM)Ifanerrorisdetectedbytheself-diagnosisf
<7. Inspection and Maintenance>7-3IM 11M10B01-01ENote) If an error (ERR) occurs, the heater is turned off. To clear the error, you need to
<7.InspectionandMaintenance>7-4IM 11M10B01-01E5) Remove the lead wire connector of the sensor from the PCB board. 6) Loosenthejointrin
<7.InspectionandMaintenance>7-5IM 11M10B01-01E7.3.2 InstallingSensor1) Pass a new sensor through the heater. SensorLead wire connect
<7. Inspection and Maintenance>7-6IM 11M10B01-01E4) Insert the ring 1 and holding plate 4 in this order into the sensor. Hold the sensor with
vIM 11M10B01-01E<INTRODUCTION>u After-sales WarrantynDonotmodifytheproduct.nDuringthewarrantyperiod,forrepairunderwarrantycarryor
<7. Inspection and Maintenance>7-7IM 11M10B01-01E(1) Loosen the four screws with washers B of holder A and remove holder A. (Figure 7.4)(2)
<7. Inspection and Maintenance>7-8IM 11M10B01-01E7.5 How to Replace the Line FilterThe line lter is inside the gas inlet connector on the rear
<8.Appendix>8-1IM 11M10B01-01E8. Appendix8.1 ParameterSettingsTable8.1 ParameterSettingList(CustomerSettings)Menu Parameter 1 Para
<8.Appendix>8-2IM 11M10B01-01E
CustomerMaintenanceParts ListOX400Low Concentration Zirconia Oxygen Analyzer[Style: S2]CMPL 11M10B01-02E2nd Edition : Apr. 2011 (YK)© Copyright Mar. 2
2CMPL 11M10B01-02E2nd Edition : Apr. 2011 (YK)Item Part No. Qty Description 1 K9643KH 1 Filter Assembly (Rc1/4)
iIM 11M10B01-01ERevision Information Title : Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: S2] Manual No. : IM 11M10B01-01EMay.
Blank Page
viIM 11M10B01-01E<INTRODUCTION>
iIM 11M10B01-01E<CONTENTS>Model OX400Low Concentration Zirconia Oxygen Analyzer[Style: S2]CONTENTSIM11M10B01-01E3rdEditionINTRODUCTION...