Yokogawa Electric CorporationUser’sManualYTA SeriesTemperature Transmitter(BRAIN Protocol)IM 01C50T03-01EIM 01C50T03-01E5th Edition
IM 01C50T03-01E3-33. OPERATIONADJJ:CAL SENSOR1J05:SNSR1 CLRJ07:IN TRIM MODEJ10:SNSR1 ZEROJ20:SNSR1 GAINJ30:SNSR1 SERIALJ60:SELF CHECKK:CAL SENSOR2K05
IM 01C50T03-01E3-43. OPERATION3.2 Setting Parameters3.2.1 Sensor ConfigurationWhen the sensor type or the number of wire connec-tions changes, the fo
IM 01C50T03-01E3-53. OPERATIONDATA DIAG PRNT ESCPARAM D10:SENSOR1 TYPE Pt200 (IEC751) D20:SENSOR WIRE 3 WIRE D40:SENSOR1 TEMP 23.56 deg
IM 01C50T03-01E3-63. OPERATION3.2.3 Unit SettingB11: PV UNIT, B21: SV UNIT, B31: TV UNIT, B41: 4VUNITSelect the engineering unit for the process v
IM 01C50T03-01E3-73. OPERATIONMemoO10: MEMO1, O20: MEMO2Up to sixteen alphanumeric characters can be entered.DescriptionO30: DESCRIPTORUp to sixteen
IM 01C50T03-01E3-83. OPERATIONmA : Displays output value in mA%: Displays output value in %mA, % : Displays output value in mA and % alternatelyINHI
IM 01C50T03-01E3-93. OPERATION(f) Display error codeM55: Err-NO DISPIf an error occurs, the error code is displayed on theLCD indicator.SHOW : Error
IM 01C50T03-01E3-103. OPERATION3.2.9 Burn Out Function(a) Sensor burn outConfigure the burn out mode in the case of sensorfailure or disconnection. W
IM 01C50T03-01E3-113. OPERATIONtransmitter will send the error code for Sensor2 failureto the integral indicator and the BT200, and outputs the“Senso
IM 01C50T03-01E3-123. OPERATIONPin position of SW2Write Protect Status YNNo F0355.EPSW PYesYNW PSW2SW2SW2Hardware Write ProtectSW1Burnout output dire
iIM 01C50T03-01ECONTENTSFD No. IM 01C50T03-01E5th Edition: Sep. 2006 (KP)All Rights Reserved, Copyright © 1998, Yokogawa Electric CorporationCONTENTS1
IM 01C50T03-01E3-133. OPERATIONTable 3.1 Zero and Gain point value for SensortrimT0301.EPSSensor typeSetting valueTC B,R,S,T +0[mV] +25[mV] E,J,K,N
IM 01C50T03-01E3-143. OPERATIONUSER CAL IGNORE : Ignore user trim valueand return to thefactory set value.USER CAL ACT : Use user trim value.NOTERega
IM 01C50T03-01E4-14. SELF-DIAGNOSTICS4. SELF-DIAGNOSTICSThe temperature transmitter continually monitors itsown performance during normal operation.
IM 01C50T03-01E4-24. SELF-DIAGNOSTICSTable 4.1 List of Error CodesIndicatorBT200 display Cause Output operation upon error ActionT0401.EPSEr-01 Outpu
IM 01C50T03-01E4-34. SELF-DIAGNOSTICS4.2 Warning1) Warning and contentsThe YTA series has a warning display function.The warning display function dis
IM 01C50T03-01E4-44. SELF-DIAGNOSTICSTable 4.2 List of warningsThe LRV setting is lower than the temperature range (GS stated value).The LRV setting
IM 01C50T03-01E4-54. SELF-DIAGNOSTICS4.3 Logging FunctionThe YTA series has the capability to store usefulinformation for trouble shooting.4.3.1 Erro
IM 01C50T03-01E5-15. LIST OF PARAMETERS5. LIST OF PARAMETERSNO. Parameter name Content R/W RemarksApplicable modelUPLOAD DATA110 310 320 SET ALLData i
IM 01C50T03-01E5-25. LIST OF PARAMETERSE SET SENSOR2 Set Sensor2 RE10SENSOR2 TYPESensor2 sensor type W TYPE B, E, J, K, L, N, R, S, T, U, W3, W5, Pt10
IM 01C50T03-01E5-35. LIST OF PARAMETERSJ CAL SENSOR1 Sensor1 sensor trim RJ05 SNSR1 CLR Sensor1 trim value clear W USER CAL ACT, USER CAL IGNOR
iiIM 01C50T03-01ECONTENTSAPPENDIX A. OPERATION OF BRAIN TERMINAL BT200... A-1A.1 Operation Key Arrangement ...
IM 01C50T03-01EA-1APPENDIX A. OPERATION OF BRAIN TERMINAL BT200APPENDIX A. OPERATION OF BRAINTERMINAL BT200A.1 Operation Key ArrangementFigure 5.3 sh
IM 01C50T03-01EA-2APPENDIX A. OPERATION OF BRAIN TERMINAL BT200A.2 Function of Operation KeysA.2.1 Entry of Alphanumeric CharactersNumbers, codes, an
IM 01C50T03-01EA-3APPENDIX A. OPERATION OF BRAIN TERMINAL BT200A.2.2 Function KeysThe functions of the function keys vary with thecommands being disp
IM 01C50T03-01EA-4APPENDIX A. OPERATION OF BRAIN TERMINAL BT200A.3 Calling of Menu AddressDATA DIAG PRNT ESCPARAM A60:SELF CHECK GOODCODE CAPS
IM 01C50T03-01EA-5APPENDIX A. OPERATION OF BRAIN TERMINAL BT200A.3.1 Data Display with BT200The following procedure is used to display data on theBT2
IM 01C50T03-01EA-6APPENDIX A. OPERATION OF BRAIN TERMINAL BT200[SETUP SCREEN][SETUP SCREEN][CONFIRMATIONSCREEN][SETTING COMPLETEDSCREEN][PARAMETER SC
IM 01C50T03-01EB-1APPENDIX B. THE SENSOR MATCHING FUNCTIONAPPENDIX B. THE SENSOR MATCHINGFUNCTIONB.1 SpecificationsFunction: The sensor-specific cons
IM 01C50T03-01EB-2APPENDIX B. THE SENSOR MATCHING FUNCTIONB.2 Operations (The SensorMatching Function)IMPORTANTThis function is effective only in thr
IM 01C50T03-01EC-1APPENDIX C. SAFETY INSTRUMENTED SYSTEMS INSTALLATIONAPPENDIX C. SAFETY INSTRUMENTEDSYSTEMS INSTALLATIONWARNINGThe contents of this
IM 01C50T03-01EC-2APPENDIX C. SAFETY INSTRUMENTED SYSTEMS INSTALLATION Table C.2.5 Proof TestingTA0102.EPSTesting method Tools required E
IM 01C50T03-01E1-11. INTRODUCTION1. INTRODUCTIONThank you for purchasing the YTA series TemperatureTransmitter.The YTA temperature transmitter is ful
IM 01C50T03-01EC-3APPENDIX C. SAFETY INSTRUMENTED SYSTEMS INSTALLATIONC.2.9 Reliability DataA detailed Failure Mode, Effects, and DiagnosticsAnalysis
Revision Record• Manual No. : IM 01C50T03-01E• Title : YTA series Temperature Transmitter (BRAIN Protocol)Edition Date Page Revised item1st Sep. 1998
IM 01C50T03-01E1-21. INTRODUCTION䊏 For Safe Use of ProductFor the protection and safety of the operator and theinstrument or the system including the
IM 01C50T03-01E1-31. INTRODUCTION䊏 ATEX DocumentationThis procedure is only applicable to the countries inEuropean Union.GBAll instruction manuals fo
IM 01C50T03-01E2-12. CONDITIONS OF COMMUNICATION LINE2. CONDITIONS OF COMMUNICATIONLINE2.1 Interconnection Between YTAand BT200WARNINGDo not attempt
IM 01C50T03-01E3-13. OPERATION3. OPERATIONIMPORTANTDo not turn off the power to the transmitterimmediately after setting the data using theBT200. If
IM 01C50T03-01E3-23. OPERATION• Sensor Backup Function (YTA320 only)See Page 3-10Configure the transmitter to automatically transferthe input from Se