User’sManualModels DX1002/DX1004/DX1006/DX1012/DX1002N/DX1004N/DX1006N/DX1012NDaqstation DX1000/DX1000NOperation GuideIM 04L41B01-02E7th Edition
8IM 04L41B01-02EIntroduction to FunctionsMeasured ItemsYou can connect DC voltage, thermocouple, RTD, and ON/OFF input and measure various values such
9IM 04L41B01-02EDX System ConfigurationThe DX can be used to configure a system as shown below.Referenced sections are of the DX1000/DX1000N User’s Ma
10IM 04L41B01-02ENames of PartsFront ViewUSB CF CARDNO.Do not remove the cover.Repair by trained personnel only.START USER FUNC ESC MENUSTOPKey panel
11IM 04L41B01-02ERear PanelEthernet portA 10Base-T port.USB port (/USB1 option)A USB Rev. 1.1 compliant port.Optional terminals (/A[ ], /F1, /R1, /TPS
12IM 04L41B01-02EDX1000/DX1000N WorkflowWhen using the DX for the first time, carry out the following procedure. InstallationWiring Power ON Environ
13IM 04L41B01-02ETurning the Power ON/OFF (DX1002, DX1004, DX1006, and DX1012 Only)Turning the Power ON CAUTIONBefore turning ON the power switch, ch
14IM 04L41B01-02EBasic OperationPanel KeysSTART USER FUNC ESC MENUSTOP START USER FUNC ESC MENUSTOP LCDKey panel Soft keysSelects the menu that is d
15IM 04L41B01-02EDisplay on the Status Display SectionThe following information is displayed in the status display section.Memory sampling statusError
16IM 04L41B01-02ERun ModesHold down FUNCfor 3 s orBasic Setting Mode> DISP/ENTER(release number 3or later)End menu > DISP/ENTER orESC > DISP/
17IM 04L41B01-02EEntering Values and CharactersThe character/number input window and DISP/ENTER key are used to set the date/time, set the display spa
PRS 108-02EUser RegistrationThank you for purchasing YOKOGAWA products.We invite you to register your products in order to receive the most up to date
18IM 04L41B01-02EThe contents of the screens in the operation examples below may change depending on which options are installed and how the settings
19IM 04L41B01-02E7. Press ESC twice or MENU once to return to the operation mode screen. On a DX with advanced security (/AS1 option; release number
20IM 04L41B01-02E5. Press DISP/ENTER once. Select First-CH.6. Press the +1 soft key once to set First-CH and Last-CH to 2.7. Press the down
21IM 04L41B01-02E14. Press DISP/ENTER once. The changed items are entered, and the cursor returns to First-CH. The changed items change from yellow to
22IM 04L41B01-02EOperation Example in the Basic Setting Mode: Changing the Scan IntervalIn this example, we will change the scan interval. Here, the s
23IM 04L41B01-02E4. Press the down arrow key twice to select A/D, Memory.5. Press DISP/ENTER once. 6. Press the down arrow key once to move
24IM 04L41B01-02EInserting/Removing a CF CardInserting a CF Card1. Open the operation cover. CF cardWith the label “This side up” facing upTh
25IM 04L41B01-02E5. Press the CF card eject button. When you eject the CF card, the CF card icon disappears. Eject buttonPress the eject butt
26IM 04L41B01-02ESaving the Setup DataIn this example, we will save the setup data to a file named “SF2” on the CF card.1. Display the operation mode
27IM 04L41B01-02ELoading the Setup DataIn this example, we will load the setup data “SF2” from the CF card and update the DX settings.This procedure o
1IM 04L41B01-02EContentsSafety Precautions ...
28IM 04L41B01-02ESetting the Input Range and AlarmSetup Example 1: Temperature Measurement Channel Type T thermocoupleChannel 10.0 to 200.0°CDXSet
29IM 04L41B01-02ESetting the Input Range and AlarmSetup Example 2: Flow Rate Measurement Channel and Alarm DXFlowmeter Channel 24 - 20 mAConvert t
30IM 04L41B01-02ESetting the DisplaySetup Example 3: Assigning Channels to GroupsIn this example, we will assign channels 1 and 2 to group 1. Chan
31IM 04L41B01-02ESetting the DisplaySetup Example 4: Setting the Time ScaleSet the time per division of the trend waveform to 2 minutes. The sampl
32IM 04L41B01-02ESetting the Data StorageSetup Example 5: Continuously Record Measured Data and Automatically SaveIn this example, we will continuousl
33IM 04L41B01-02E(2) Method of Storing to the CF Card (Auto Save ON/OFF) Select the Environment tab > Security, Media save 3(3) Save the Setti
34IM 04L41B01-02E(7) Save Destination Directory (within the CF Card) Select the Menu tab > Data save > Save directory. 6Operation complete.
35IM 04L41B01-02ESetup Example 6: Saving Measured Data at the Specified TimeUsing the settings of Setup Example 5, we will save the measured data once
36IM 04L41B01-02ECustomizing the OperationSetup Example 7: Assigning the Screen Image Data Storage Function to the USER KeyIn this example, we will se
37IM 04L41B01-02ECustomizing the OperationSetup Example 8: Registering Frequently Used Screens to the Favorite KeyUp to eight operation mode screens t
2IM 04L41B01-02EContentsDaqstation DX1000/DX1000N User’s Manual (Electronic Manual Provided on the Accompanying CD)Chapter 1 Overview of FunctionsCha
38IM 04L41B01-02E5. Press the favorite number (1 to 8) soft key.6. Press the Regist soft key. Show the window for entering the display name.
39IM 04L41B01-02EOperationStarting Memory Sampling1. Press START once.On a DX with advanced security (/AS1 option; release number 4 or later) display
40IM 04L41B01-02ESwitching the Trend Display, Digital Display, and Bar Graph Display1. Press DISP/ENTER once to show the display selection menu.
41IM 04L41B01-02EOperationWriting the Message “START”Registering the Word “START” in Message Number 1On a DX with advanced security (/AS1 option; rele
42IM 04L41B01-02EWriting Message Number 1 “START”This operation can only be carried out while memory sampling is in progress. The message is displayed
43IM 04L41B01-02EConnecting to an Ethernet NetworkSetup Example 9: Monitoring the DX on a PC BrowserIn this example, we will connect the PC and the DX
44IM 04L41B01-02E(2) Enabling the Web Server Function on the DX Select the Menu tab > Communication (Ethernet) > Server > Server. 2 (3)
45IM 04L41B01-02E(6) Checking the Connection Send the command below from the PC and check that a correct response is returned. Send >ping 192.16
46IM 04L41B01-02ESetup Example 10: Automatically Transferring the Measured Data File to an FTP ServerIn this example, we will configure the DX so that
47IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default Values(3) Data to Be Transferred to the
3IM 04L41B01-02EThank you for purchasing the Daqstation DX1000/DX1000N (DX).This manual describes the basic operating procedures and installation and
48IM 04L41B01-02EUsing DAQSTANDARDDisplaying the Measured Data on DAQSTANDARDIn this example, we will display the measured data using the accompanying
49IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesInstallation and WiringInstallatio
50IM 04L41B01-02EInstallation ProcedureInstallation Procedure (Panel Mount Type)Use a steel panel of thickness 2 mm to 26 mm.1. Insert DX from the fr
51IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesExternal Dimensions and Panel Cut
52IM 04L41B01-02EInput Signal Wiring WARNING• To prevent electric shock while wiring, ensure that the power supply source is turned OFF.CAUTION•
53IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesWiring ProcedureA terminal cover i
54IM 04L41B01-02EWiring Screw Terminals–++–DC voltage inputTC inputDC voltage input/DI (ON/OFF) inputRTD inputDC current inputDC current inputShunt re
55IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesOptional Terminal Wiring WARNING•
56IM 04L41B01-02EArrangement of the Optional Terminals Optional terminal block02 0104 03NCCNONCCNONCCNONCCNONCCNONCCNONCCNONCCNONCCNONCCNONCCNONCCNO
57IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesNCCNONCCNOPulseinputPulseinput02 0
4IM 04L41B01-02ESafety PrecautionsThis instrument conforms to IEC safety class I (provided with terminal for protective grounding), Installation Categ
58IM 04L41B01-02EAlarm Output Terminal (/A1, /A2, and /A3), FAIL Output Terminal, and Status Output Terminal (/F1)CNCNOOutput format: Relay contactCo
59IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesConnecting to the RS-422/485 (/C3)
60IM 04L41B01-02EPower Supply WiringPanel Mount Type, or Desktop Type with /P1 (Models with /H5 and /P1 Options)Precautions to Be Taken While Wiring t
61IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesDesktop Type (/H5[ ] Option)Precau
62IM 04L41B01-02ERecommended Replacement Periods for Worn PartsTo preserve the reliability of the DX and to use the DX in a good condition for an exte
63IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesSetup Items and Default ValuesThe
64IM 04L41B01-02EBasic setting mode menu Page 77Page 77Page 77Page 77Page 77*1Page 78Page 78Page 78Page 78*1Page 79*1Page 80Page 80Page 80Page 81*
65IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesSetup Items in Setting Mode and Th
66IM 04L41B01-02ESetup Item Selectable Range or Selections Default Value SettingMode=DIRangeSpan LowerSpan UpperLevel/Cont0, 10, 1Level01Mode=1-5VRang
67IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesMeas channel > ColorSetup Item
5IM 04L41B01-02E• Copying the software for any purposes other than backup is strictly prohibited.• Please store the original media containing the so
68IM 04L41B01-02EMath channel > Expression, AlarmSetup Item Selectable Range or Selections Default Value SettingFirst-CH, Last-CH Channel number 10
69IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesMath channel > PartialSetup ite
70IM 04L41B01-02EDisplay > Trend/Save intervalSetup Item Selectable Range or Selections Default Value SettingTrend interval [/div] Save interval S
71IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesGroup set, Trip lineSetup Item Sel
72IM 04L41B01-02ETimer, Event action > Event actionSetup Item Selectable Range or Selections Default Value SettingLogic box number 1/2/3/.../39/40
73IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesSetup Item Selectable Range or Sel
74IM 04L41B01-02EData save > File header, File nameSetup Item Selectable Range or Selections Default Value SettingFile header > Characters 50 ch
75IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesMulti batch tab*2Group set, Trip l
76IM 04L41B01-02ELoad event dataSetup Item Selectable Range or Selections Default Value SettingKind CF/USB CFLoad settingsSetup Item Selectable Range
77IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesSetup Items in Basic Setting Mode
6IM 04L41B01-02E Standard AccessoriesThe standard accessories below are supplied with the instrument. Check that all contents are present and undamag
78IM 04L41B01-02ESecurity, Media saveSetup Item Selectable Range or Selections Default Value SettingWithout /AS1Security > KeySecurity > Communi
79IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesReportSetup Item Selectable Range
80IM 04L41B01-02EMenu tab*2Alarm > Basic settingsSetup Item Selectable Range or Selections Default Value SettingReflashRate of change > Decrease
81IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesKeylock > Password, Key action,
82IM 04L41B01-02ELogin > User settingsSetup Item Selectable Range or Selections Default Value SettingWithout /AS1User number 1 to 30 1ModeUser name
83IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesReport > Basic settingsSetup It
84IM 04L41B01-02ECommunication (Ethernet) > IP-addressSetup Item Selectable Range or Selections Default Value SettingDHCP Use/Not NotDNS accession
85IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesSetup Items and Default ValuesComm
86IM 04L41B01-02ECommunication (Ethernet) > E-Mail > Scheduled settingsSetup Item Selectable Range or Selections Default Value SettingRecipient
87IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesCommunication (Ethernet) > FTP
7IM 04L41B01-02EProtection of Environment Control of Pollution Caused by the ProductFor details, see the Control of Pollution Caused by the Product (
88IM 04L41B01-02ECommunication (Ethernet) > Password management > Certification key (/AS1 option)*3Setup Item Selectable Range or Selections Def
89IM 04L41B01-02EInstallation and WiringRecommended Replacement Periods for Worn PartsSetup Items and Default ValuesStatus relay*3Setup Item Selectabl
90IM 04L41B01-02EFile/Initialize tab*2Load settingsSetup Item Selectable Range or Selections Default Value SettingWithout /AS1Load settings > Kind