User’sManualYokogawa Electric CorporationBRAIN TERMINALModel BT200BRAIN TERMINALIM 01C00A11-01E5th Edition
<Toc> <Ind> <2. INTRODUCTION TO BT200>2-3IM 01C00A11-01E2.2 SPECIFICATIONEquipment SpecificationsApplicable EquipmentDPharp SeriesAD
2-4<Toc> <Ind> <2. INTRODUCTION TO BT200>IM 01C00A11-01E• Automatic Power-off ··· The terminal switches offautomatically if no key a
<Toc> <Ind> <2. INTRODUCTION TO BT200>2-5IM 01C00A11-01E2.3 COMPONENT NAMEFuigure 2.3 Component Name
2-6<Toc> <Ind> <2. INTRODUCTION TO BT200>IM 01C00A11-01E2.4 DIMENSIONSFuigure 2.4 Dimensions
<Toc> <Ind> <3. CONNECTION >3-1IM 01C00A11-01E3. CONNECTION3.1 PLUGGING THE CABLEINTO THE BT200When plugging the dedicated communic
3-2<Toc> <Ind> <3. CONNECTION >IM 01C00A11-01E3.2 CONNECTION WITH BRAINSERIES INSTRUMENTS(1) CONNECTION IN THE INSTRUMENTTERMINAL BO
<Toc> <Ind> <3. CONNECTION >3-3IM 01C00A11-01E(3) CONNECTION TO SIGNAL CONDITIONERESC (signal conditioner communication card) orEXT
<Toc> <Ind> <4. BT200 BASIC OPERATIONS >4-1IM 01C00A11-01E4. BT200 BASIC OPERATIONS4.1 KEY LAYOUT AND DISPLAYF0401.EPSLCD(21 chara
4-2<Toc> <Ind> <4. BT200 BASIC OPERATIONS >IM 01C00A11-01E4.2 KEY DESCRIPTIONS(1) ALPHANUMERIC KEYS AND SHIFT KEYSYou can use the al
<Toc> <Ind> <4. BT200 BASIC OPERATIONS >4-3IM 01C00A11-01E*Use the function key [F2] CAPS to select betweenuppercase and lowercase
<Toc> <Ind> <CONTENTS>iIM 01C00A11-01ECONTENTS5. USING THE BT200 FUNCTIONS ...5-15.1 SETTING UP DATA IN A BATCH(UPLOAD
4-4<Toc> <Ind> <4. BT200 BASIC OPERATIONS >IM 01C00A11-01E(2) FUNCTION KEYSThe functions of the function keys depend on thefunction
<Toc> <Ind> <4. BT200 BASIC OPERATIONS >4-5IM 01C00A11-01E4.3 BT200 FUNCTION CONFIGURATION––WELCOME––BRAIN TERMINAL ID: BT200
4-6<Toc> <Ind> <4. BT200 BASIC OPERATIONS >IM 01C00A11-01E4.4 BASIC OPERATIONSThis section illustrates the basic operations ofthe BT
<Toc> <Ind> <4. BT200 BASIC OPERATIONS >4-7IM 01C00A11-01EPARAM 01:MODEL EJ110-DM 02:TAG NO. TABC-101 03:SELF CHECK
4-8<Toc> <Ind> <4. BT200 BASIC OPERATIONS >IM 01C00A11-01E(2) DISPLAY PARAMETERSa) Menu panel operationsThe menu panel contains a li
<Toc> <Ind> <4. BT200 BASIC OPERATIONS >4-9IM 01C00A11-01Eb) Call the parameter panelOn initial data panel, type [F4] OK or[ENT] to
4-10<Toc> <Ind> <4. BT200 BASIC OPERATIONS >IM 01C00A11-01EHelpful hint 1• Typing letters in a panel allows you to make a directsele
<Toc> <Ind> <4. BT200 BASIC OPERATIONS >4-11IM 01C00A11-01EK10 ESCF0415-2.EPSPARAM K10:OUTPUT X % 0.0 % K60:SELF CHECK GO
4-12<Toc> <Ind> <4. BT200 BASIC OPERATIONS >IM 01C00A11-01E(3) CHANGE SETUP DATATo change setup data, select a parameter to changefr
<Toc> <Ind> <4. BT200 BASIC OPERATIONS >4-13IM 01C00A11-01EPARAM C10:TAG NO. TABC-101 C20:PRESS UNIT mmH2O C21:LRV0
<Toc> <Ind> <INTRODUCTION>1IM 01C00A11-01EINTRODUCTIONThank you for choosing the BT200 BRAIN TERMI-NAL.This user’s manual describes
4-14<Toc> <Ind> <4. BT200 BASIC OPERATIONS >IM 01C00A11-01E(4) ENTER A SECURITY CODE (IF ONE HASBEEN SET)To alter setup data for the
<Toc> <Ind> <5. USING THE BT200 FUNCTIONS >5-1IM 01C00A11-01E5. USING THE BT200FUNCTIONS5.1 SETTING UP DATA IN ABATCH (UPLOAD/DOWNLO
5-2<Toc> <Ind> <5. USING THE BT200 FUNCTIONS >IM 01C00A11-01EF0501-2.EPSUP LOAD MODEL EJ110-DM UPLOAD start? push ENTER key
<Toc> <Ind> <5. USING THE BT200 FUNCTIONS >5-3IM 01C00A11-01EF0502-2.EPSDNLOAD MODEL EJ110-DM DOWNLOAD start? push ENTER key
5-4<Toc> <Ind> <5. USING THE BT200 FUNCTIONS >IM 01C00A11-01E5.2 USING PRINTER FUNCTIONS(BT200-P00)The BT200-P00 BRAIN TERMINAL with
<Toc> <Ind> <5. USING THE BT200 FUNCTIONS >5-5IM 01C00A11-01E(2) PRINTING DISPLAY PARAMETERSPARAM A10:OUTPUT 100.0 % A11:ENGR. O
5-6<Toc> <Ind> <5. USING THE BT200 FUNCTIONS >IM 01C00A11-01EF0506-2.EPSPRINT menu data printing please wait... F4:STOP printin
<Toc> <Ind> <5. USING THE BT200 FUNCTIONS >5-7IM 01C00A11-01E(5) PRINTING UPLOADED OR DOWNLOADEDDATAa) UPLOADUPLOAD MODEL EJ110
5-8<Toc> <Ind> <5. USING THE BT200 FUNCTIONS >IM 01C00A11-01E(6) PRINTING A LIST OF SELF-CHECK INFOR-MATIONPARAM C10:TAG NO. T
<Toc> <Ind> <5. USING THE BT200 FUNCTIONS >5-9IM 01C00A11-01E(1) ID (Identification Code) SETUPEach BT200 BRAIN TERMINAL is assigned
<Toc> <Ind> <1. PRECAUTIONS>1-1IM 01C00A11-01E1. PRECAUTIONS1.1 CHECK THE CONTENTS OFTHE PACKAGEThe BT200 BRAIN TERMINAL are thoroug
5-10<Toc> <Ind> <5. USING THE BT200 FUNCTIONS >IM 01C00A11-01Eb) Japanese to EnglishF0515-1.EPS1.Press [F1] to select the uni
<Toc> <Ind> <5. USING THE BT200 FUNCTIONS >5-11IM 01C00A11-01E(3) LCD CONTRAST CONTROLThe BT200 features software capability to cont
<Toc> <Ind> <6. MAINTENANCE>6-1IM 01C00A11-01E6. MAINTENANCE6.1 REPLACING BATTERIESWhen the BATTERY indicator is lit on the screen,
6-2<Toc> <Ind> <6. MAINTENANCE >IM 01C00A11-01EManufacture Model Type VoltageDURACELL MN1500 Alkaline- 1.5 V(PC1500) manganeseFigure
<Toc> <Ind> <6. MAINTENANCE>6-3IM 01C00A11-01E6.2 LOADING ROLL PAPERWith the BT200-P00 BRAIN TERMINAL with aprinter, follow these st
<Toc> <Ind> <7. TROUBLESHOOTING>7-1IM 01C00A11-01E7. TROUBLESHOOTINGHere are simple troubleshooting tips. If problemspersist after a
7-2<Toc> <Ind> <7. TROUBLESHOOTING>IM 01C00A11-01E3. THE BATTERY INDICATOR IS ONThe BATTERY indicator, when on, warns you of a lowba
<Toc> <Ind> <7. TROUBLESHOOTING>7-3IM 01C00A11-01EFigure 7.2 Connector pin numbersFigure 7.3 Wiring
<Toc> <Ind> <APPENDIX>A-1IM 01C00A11-01EAPPENDIX A1. Setting and Changing SECURITY CODEThe BT200 is shipped without a security code
A-2<Toc> <Ind> <APPENDIX>IM 01C00A11-01EB) CHANGING THE SECURITY CODETo change a security code after it has been regis-tered, follow
1-2<Toc> <Ind> <1. PRECAUTIONS>IM 01C00A11-01E1.2 PRECAUTIONS IN HANDLING(1) WHERE TO STOREStore the unit in places meeting the foll
<Toc> <Ind> <APPENDIX>A-3IM 01C00A11-01E3. When You Forget the SECURITY CODEWhen you forget the security code that has beenregistere
A-4<Toc> <Ind> <APPENDIX>IM 01C00A11-01EAPPENDIX BInstallation of CSA Intrinsically Safe TypeBT200 BRAIN TERMINAL (Model BT200-N00/C
<Toc> <Ind> <APPENDIX>A-5IM 01C00A11-01E• Safety Barriers Parameters (CSA Certified Barriers)Voc 28VIsc 143mAPmax 889mW• Intri
<Toc> <Ind> <REVISION RECORD>R-1IM 01C00A11-01ERev1.EPSEdition DatePageRevised ItemREVISION RECORD1st1994May2ndNew Publication·Add S
<Toc> <Ind> <1. PRECAUTIONS>1-3IM 01C00A11-01Ed. Following communication is not permitted as itmay result communication error or ind
1-4<Toc> <Ind> <1. PRECAUTIONS>IM 01C00A11-01E(5) EMC CONFORMITY STANDARDS, EN61326, AS/NZS CISPR11
<Toc> <Ind> <2. INTRODUCTION TO BT200>2-1IM 01C00A11-01E2. INTRODUCTION TOBT200The BT200 BRAIN TERMINAL is a portable terminalused i
2-2<Toc> <Ind> <2. INTRODUCTION TO BT200>IM 01C00A11-01E2.1 WHAT THE BT200 CAN DOThe BT200 offers the following exclusive features:F